A Microwave Metamaterial Inspired Sensor for Non-Invasive Blood Glucose Monitoring
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21460%2F15%3A00235288" target="_blank" >RIV/68407700:21460/15:00235288 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.13164/re.2015.0877" target="_blank" >http://dx.doi.org/10.13164/re.2015.0877</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.13164/re.2015.0877" target="_blank" >10.13164/re.2015.0877</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
A Microwave Metamaterial Inspired Sensor for Non-Invasive Blood Glucose Monitoring
Popis výsledku v původním jazyce
In this paper, a microwave sensor based on an artificial transmission line is proposed for non-invasive blood glucose monitoring. A corresponding numerical model of the sensor implemented in microstrip technology is created in the commercial full-wave numerical simulation tool COMSOL Multiphysics and virtually tested by means of numerical simulations. Blood-glucose solution models with various blood glucose concentrations are used as a model of a biological tissue under test. Furthermore, a possible methodology for performing non-invasive tests is proposed. Sensitivity of the sensor developed here is compared to a sensor based on a section of a conventional microstrip transmission line of the same length and width.
Název v anglickém jazyce
A Microwave Metamaterial Inspired Sensor for Non-Invasive Blood Glucose Monitoring
Popis výsledku anglicky
In this paper, a microwave sensor based on an artificial transmission line is proposed for non-invasive blood glucose monitoring. A corresponding numerical model of the sensor implemented in microstrip technology is created in the commercial full-wave numerical simulation tool COMSOL Multiphysics and virtually tested by means of numerical simulations. Blood-glucose solution models with various blood glucose concentrations are used as a model of a biological tissue under test. Furthermore, a possible methodology for performing non-invasive tests is proposed. Sensitivity of the sensor developed here is compared to a sensor based on a section of a conventional microstrip transmission line of the same length and width.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
JA - Elektronika a optoelektronika, elektrotechnika
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2015
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Radioengineering
ISSN
1210-2512
e-ISSN
—
Svazek periodika
2015
Číslo periodika v rámci svazku
4
Stát vydavatele periodika
CZ - Česká republika
Počet stran výsledku
8
Strana od-do
877-884
Kód UT WoS článku
000366800200001
EID výsledku v databázi Scopus
2-s2.0-84951025335