FITPix - fast interface for Timepix pixel detectors
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21670%2F11%3A00175134" target="_blank" >RIV/68407700:21670/11:00175134 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1088/1748-0221/6/01/C01079" target="_blank" >http://dx.doi.org/10.1088/1748-0221/6/01/C01079</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1088/1748-0221/6/01/C01079" target="_blank" >10.1088/1748-0221/6/01/C01079</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
FITPix - fast interface for Timepix pixel detectors
Popis výsledku v původním jazyce
The semiconductor pixel detector Timepix contains an array of 256 x 256 square pixels with pitch 55 ?m. In addition to high spatial granularity the single quantum counting detector Timepix can provide also energy or time information in each pixel. This device is a powerful tool for radiation and particle detection, imaging and tracking. A new readout interface for silicon pixel detectors of the Medipix family has been developed in order to provide a higher frame rate and enhanced flexibility of operation. The interface consists of a field programmable gate array, a USB 2.0 interface chip, DAC, ADC and a sensor bias voltage generator. The entire system is fully powered through the USB bus. FITPix supports also readout from several detectors in chain inwhich case just an external power source is required. FITPix is a fully flexible device and the user needs no other equipment. FITPix combines high performance and mobility and it opens new fields of applications.
Název v anglickém jazyce
FITPix - fast interface for Timepix pixel detectors
Popis výsledku anglicky
The semiconductor pixel detector Timepix contains an array of 256 x 256 square pixels with pitch 55 ?m. In addition to high spatial granularity the single quantum counting detector Timepix can provide also energy or time information in each pixel. This device is a powerful tool for radiation and particle detection, imaging and tracking. A new readout interface for silicon pixel detectors of the Medipix family has been developed in order to provide a higher frame rate and enhanced flexibility of operation. The interface consists of a field programmable gate array, a USB 2.0 interface chip, DAC, ADC and a sensor bias voltage generator. The entire system is fully powered through the USB bus. FITPix supports also readout from several detectors in chain inwhich case just an external power source is required. FITPix is a fully flexible device and the user needs no other equipment. FITPix combines high performance and mobility and it opens new fields of applications.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
BG - Jaderná, atomová a molekulová fyzika, urychlovače
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/LC06041" target="_blank" >LC06041: Příprava, modifikace a charakterizace materiálů energetickým zářením</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2011
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Journal of Instrumentation
ISSN
1748-0221
e-ISSN
—
Svazek periodika
6
Číslo periodika v rámci svazku
—
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
6
Strana od-do
1-6
Kód UT WoS článku
000291345600084
EID výsledku v databázi Scopus
—