Analogue signal from common electrode of pixelated detector for triggering and spectroscopy
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21670%2F11%3A00188105" target="_blank" >RIV/68407700:21670/11:00188105 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1088/1748-0221/6/11/C11023" target="_blank" >http://dx.doi.org/10.1088/1748-0221/6/11/C11023</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1088/1748-0221/6/11/C11023" target="_blank" >10.1088/1748-0221/6/11/C11023</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Analogue signal from common electrode of pixelated detector for triggering and spectroscopy
Popis výsledku v původním jazyce
A significant advantage of novel high granularity and highly sensitive semiconductor pixel detectors is the possibility toof directly observeation of single tracks of charged particles. In many cases, however, these particles are accompanied with unwanted background radiation overlapping traces of interest. The detection selectivity can be increased using a triggering approach usually provided by an external trigger from other detecting devices such as ionization chambers, scintillating or semiconductordetectors. A self-trigger from the same sensor would be highly desirable. Unfortunately the Medipix/Timepix devices are not equipped with such self-trigger feature. A solution which is discussed presented in this contribution makes uses of the analog signal taken from the common electrode of the pixelated sensor. This signal, called back-side-pulse, is amplified by a custom made charge sensitive preamplifier whichand, after shaping, can be used as fast trigger and as independent spectro
Název v anglickém jazyce
Analogue signal from common electrode of pixelated detector for triggering and spectroscopy
Popis výsledku anglicky
A significant advantage of novel high granularity and highly sensitive semiconductor pixel detectors is the possibility toof directly observeation of single tracks of charged particles. In many cases, however, these particles are accompanied with unwanted background radiation overlapping traces of interest. The detection selectivity can be increased using a triggering approach usually provided by an external trigger from other detecting devices such as ionization chambers, scintillating or semiconductordetectors. A self-trigger from the same sensor would be highly desirable. Unfortunately the Medipix/Timepix devices are not equipped with such self-trigger feature. A solution which is discussed presented in this contribution makes uses of the analog signal taken from the common electrode of the pixelated sensor. This signal, called back-side-pulse, is amplified by a custom made charge sensitive preamplifier whichand, after shaping, can be used as fast trigger and as independent spectro
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
BG - Jaderná, atomová a molekulová fyzika, urychlovače
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>Z - Vyzkumny zamer (s odkazem do CEZ)<br>S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2011
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Journal of Instrumentation
ISSN
1748-0221
e-ISSN
—
Svazek periodika
6
Číslo periodika v rámci svazku
—
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
6
Strana od-do
1-6
Kód UT WoS článku
000298320400023
EID výsledku v databázi Scopus
—