X-ray color imaging with 3D sensitive voxel detector
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21670%2F11%3A00188283" target="_blank" >RIV/68407700:21670/11:00188283 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1088/1748-0221/6/12/C12014" target="_blank" >http://dx.doi.org/10.1088/1748-0221/6/12/C12014</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1088/1748-0221/6/12/C12014" target="_blank" >10.1088/1748-0221/6/12/C12014</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
X-ray color imaging with 3D sensitive voxel detector
Popis výsledku v původním jazyce
X-ray imaging is today widely used in a broad range of applications. Nevertheless some limitations are represented by the inability to distinguish between a thick layer of low Z material and a thin layer of high Z material, and by the beam hardening, where the incident X-ray spectrum is modified as the beam traverses the sample. Such effects cause problems in many applications (e.g. CT reconstruction) generating artifacts and worsening the spatial resolution. This work presents a new technique allowingspectral sensitivity using a new 3D voxel detector based on the Timepix pixel detector. The device is designed as a layered stack of several Timepix sensors. The readout chip is thinned down to reduce the amount of insensitive absorbing material. Every single layers in the stack act as a filter, i.e. each stack layer visualizes a different part of the spectrum attenuated by the object giving further information about the object composition. The comparison of attenuation levels observed i
Název v anglickém jazyce
X-ray color imaging with 3D sensitive voxel detector
Popis výsledku anglicky
X-ray imaging is today widely used in a broad range of applications. Nevertheless some limitations are represented by the inability to distinguish between a thick layer of low Z material and a thin layer of high Z material, and by the beam hardening, where the incident X-ray spectrum is modified as the beam traverses the sample. Such effects cause problems in many applications (e.g. CT reconstruction) generating artifacts and worsening the spatial resolution. This work presents a new technique allowingspectral sensitivity using a new 3D voxel detector based on the Timepix pixel detector. The device is designed as a layered stack of several Timepix sensors. The readout chip is thinned down to reduce the amount of insensitive absorbing material. Every single layers in the stack act as a filter, i.e. each stack layer visualizes a different part of the spectrum attenuated by the object giving further information about the object composition. The comparison of attenuation levels observed i
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
BG - Jaderná, atomová a molekulová fyzika, urychlovače
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2011
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Journal of Instrumentation
ISSN
1748-0221
e-ISSN
—
Svazek periodika
6
Číslo periodika v rámci svazku
12
Stát vydavatele periodika
IT - Italská republika
Počet stran výsledku
6
Strana od-do
"nečíslov. (ArtNo=C12014)"
Kód UT WoS článku
000299536600014
EID výsledku v databázi Scopus
—