Characterization of Medipix3 With Synchrotron Radiation
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21670%2F11%3A00191133" target="_blank" >RIV/68407700:21670/11:00191133 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1109/TNS.2010.2089062" target="_blank" >http://dx.doi.org/10.1109/TNS.2010.2089062</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1109/TNS.2010.2089062" target="_blank" >10.1109/TNS.2010.2089062</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Characterization of Medipix3 With Synchrotron Radiation
Popis výsledku v původním jazyce
Medipix3 is the latest generation of photon counting readout chips of the Medipix family. With the same dimensions as Medipix2 (256 x 256 pixels of 55 mu m x 55 mu m pitch each), Medipix3 is however implemented in an 8-layer metallization 0.13 mu m CMOStechnology which leads to an increase in the functionality associated with each pixel over Medipix2. One of the new operational modes implemented in the front-end architecture is the Charge Summing Mode (CSM). This mode consists of a charge reconstruction and hit allocation algorithm which eliminates event-by-event the low energy counts produced by charge-shared events between adjacent pixels. The present work focuses on the study of the CSM mode and compares it to the Single Pixel Mode (SPM) which is the conventional readout method for these kind of detectors and it is also implemented in Medipix3. Tests of a Medipix3 chip bump-bonded to a 300 mu m thick silicon photodiode sensor were performed at the Diamond Light Source synchrotron t
Název v anglickém jazyce
Characterization of Medipix3 With Synchrotron Radiation
Popis výsledku anglicky
Medipix3 is the latest generation of photon counting readout chips of the Medipix family. With the same dimensions as Medipix2 (256 x 256 pixels of 55 mu m x 55 mu m pitch each), Medipix3 is however implemented in an 8-layer metallization 0.13 mu m CMOStechnology which leads to an increase in the functionality associated with each pixel over Medipix2. One of the new operational modes implemented in the front-end architecture is the Charge Summing Mode (CSM). This mode consists of a charge reconstruction and hit allocation algorithm which eliminates event-by-event the low energy counts produced by charge-shared events between adjacent pixels. The present work focuses on the study of the CSM mode and compares it to the Single Pixel Mode (SPM) which is the conventional readout method for these kind of detectors and it is also implemented in Medipix3. Tests of a Medipix3 chip bump-bonded to a 300 mu m thick silicon photodiode sensor were performed at the Diamond Light Source synchrotron t
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
BG - Jaderná, atomová a molekulová fyzika, urychlovače
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2011
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
IEEE Transactions on Nuclear Science
ISSN
0018-9499
e-ISSN
—
Svazek periodika
58
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
10
Strana od-do
323-332
Kód UT WoS článku
000287086200020
EID výsledku v databázi Scopus
—