BrachyView, a novel inbody imaging system for HDR prostate brachytherapy: design and Monte Carlo feasibility study
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21670%2F13%3A00216015" target="_blank" >RIV/68407700:21670/13:00216015 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1118/1.4808360" target="_blank" >http://dx.doi.org/10.1118/1.4808360</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1118/1.4808360" target="_blank" >10.1118/1.4808360</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
BrachyView, a novel inbody imaging system for HDR prostate brachytherapy: design and Monte Carlo feasibility study
Popis výsledku v původním jazyce
High dose rate (HDR) brachytherapy is a form of radiation therapy for treating prostate cancer whereby a high activity radiation source is moved between predefined positions inside applicators inserted within the treatment volume. Accurate positioning ofthe source is essential in delivering the desired dose to the target area while avoiding radiation injury to the surrounding tissue. In this paper, HDR BrachyView, a novel inbody dosimetric imaging system for real time monitoring and verification of theradioactive seed position in HDR prostate brachytherapy treatment is introduced. The current prototype consists of a 15 x 60 mm(2) silicon pixel detector with a multipinhole tungsten collimator placed 6.5 mm above the detector. Seven identical pinholesallow full imaging coverage of the entire treatment volume. The combined pinhole and pixel sensor arrangement is geometrically designed to be able to resolve the three-dimensional location of the source. The probe may be rotated to keep t
Název v anglickém jazyce
BrachyView, a novel inbody imaging system for HDR prostate brachytherapy: design and Monte Carlo feasibility study
Popis výsledku anglicky
High dose rate (HDR) brachytherapy is a form of radiation therapy for treating prostate cancer whereby a high activity radiation source is moved between predefined positions inside applicators inserted within the treatment volume. Accurate positioning ofthe source is essential in delivering the desired dose to the target area while avoiding radiation injury to the surrounding tissue. In this paper, HDR BrachyView, a novel inbody dosimetric imaging system for real time monitoring and verification of theradioactive seed position in HDR prostate brachytherapy treatment is introduced. The current prototype consists of a 15 x 60 mm(2) silicon pixel detector with a multipinhole tungsten collimator placed 6.5 mm above the detector. Seven identical pinholesallow full imaging coverage of the entire treatment volume. The combined pinhole and pixel sensor arrangement is geometrically designed to be able to resolve the three-dimensional location of the source. The probe may be rotated to keep t
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
FS - Lékařská zařízení, přístroje a vybavení
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/TA01010164" target="_blank" >TA01010164: Pokročilé techniky detekce ionizujícího záření</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2013
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Medical Physics
ISSN
0094-2405
e-ISSN
—
Svazek periodika
40
Číslo periodika v rámci svazku
7
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
9
Strana od-do
—
Kód UT WoS článku
000321272200019
EID výsledku v databázi Scopus
—