The portable device for continual measurement of radon progenies on filter using the detector Timepix
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21670%2F15%3A00230405" target="_blank" >RIV/68407700:21670/15:00230405 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/86652052:_____/15:#0000314
Výsledek na webu
<a href="http://dx.doi.org/10.1093/rpd/ncv343" target="_blank" >http://dx.doi.org/10.1093/rpd/ncv343</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1093/rpd/ncv343" target="_blank" >10.1093/rpd/ncv343</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
The portable device for continual measurement of radon progenies on filter using the detector Timepix
Popis výsledku v původním jazyce
In this article, a portable device was presented for continual measuring of equilibrium equivalent concentration (EEC) of (222)Rn based on the Timepix detector with 300-m-thick active layer. In order to have a portable device, a filtration head was developed for collecting short-lived radon progenies attached on aerosols. The short-lived progenies are estimated from analysing alphas from decay of (218,214)Po from Millipore filter after termination of filtration. Comparison with beta measurement was done as well. The dependence of EEC on an air flow and filtration time was studied. The low-level detection limit for EEC was estimated from the last 10 min of 3-h decay measurement and was found in the range of 40-70 Bq m(-3). EEC was measured in National Radiation Protection Institute radon chamber, and results were compared with the commercial detector Fritra4.
Název v anglickém jazyce
The portable device for continual measurement of radon progenies on filter using the detector Timepix
Popis výsledku anglicky
In this article, a portable device was presented for continual measuring of equilibrium equivalent concentration (EEC) of (222)Rn based on the Timepix detector with 300-m-thick active layer. In order to have a portable device, a filtration head was developed for collecting short-lived radon progenies attached on aerosols. The short-lived progenies are estimated from analysing alphas from decay of (218,214)Po from Millipore filter after termination of filtration. Comparison with beta measurement was done as well. The dependence of EEC on an air flow and filtration time was studied. The low-level detection limit for EEC was estimated from the last 10 min of 3-h decay measurement and was found in the range of 40-70 Bq m(-3). EEC was measured in National Radiation Protection Institute radon chamber, and results were compared with the commercial detector Fritra4.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
DL - Jaderné odpady, radioaktivní znečištění a kontrola
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/TE01020445" target="_blank" >TE01020445: Centrum rozvoje technologií pro jadernou a radiační bezpečnost: RANUS - TD</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2015
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Radiation Protection Dosimetry
ISSN
0144-8420
e-ISSN
—
Svazek periodika
164
Číslo periodika v rámci svazku
4
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
4
Strana od-do
493-496
Kód UT WoS článku
000356726000008
EID výsledku v databázi Scopus
2-s2.0-84942097067