Experimental analysis of Schottky-barrier 4H-SiC-detector-response to fast neutrons: Comparison with Si and diamond detectors
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21670%2F25%3A00387485" target="_blank" >RIV/68407700:21670/25:00387485 - isvavai.cz</a>
Výsledek na webu
<a href="https://doi.org/10.1051/epjconf/202533810016" target="_blank" >https://doi.org/10.1051/epjconf/202533810016</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1051/epjconf/202533810016" target="_blank" >10.1051/epjconf/202533810016</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Experimental analysis of Schottky-barrier 4H-SiC-detector-response to fast neutrons: Comparison with Si and diamond detectors
Popis výsledku v původním jazyce
The response of 4H-SiC detectors to 17 MeV neutrons was experimentally analysed using detectors composed solely of carbon or silicon. The detectors were tested using three different neutron energies (14.5 MeV, 15 MeV and 17 MeV) at two different facilities with T(d,n) neutron generators, the Van de Graaff laboratory at IEAP CTU in Prague and at the DANAIDES facility at CEA research center in Cadarache. The calibration via interactions on 28Si nuclei, the (n, α) nuclear reaction to ground and excited states of 25Mg, was made. The standard deviation of experimental values measured by Si detector from prepared calibration curve, was of 0.073 MeV, which represents the uncertainty of less than 1% in this energy range ( 10 MeV). The experiment showed that the pulse height defect affects the measured spectra the way that the peak energies from the C interactions are shifted by about 0.5 MeV to higher values when the calibration from the Si interactions is used.
Název v anglickém jazyce
Experimental analysis of Schottky-barrier 4H-SiC-detector-response to fast neutrons: Comparison with Si and diamond detectors
Popis výsledku anglicky
The response of 4H-SiC detectors to 17 MeV neutrons was experimentally analysed using detectors composed solely of carbon or silicon. The detectors were tested using three different neutron energies (14.5 MeV, 15 MeV and 17 MeV) at two different facilities with T(d,n) neutron generators, the Van de Graaff laboratory at IEAP CTU in Prague and at the DANAIDES facility at CEA research center in Cadarache. The calibration via interactions on 28Si nuclei, the (n, α) nuclear reaction to ground and excited states of 25Mg, was made. The standard deviation of experimental values measured by Si detector from prepared calibration curve, was of 0.073 MeV, which represents the uncertainty of less than 1% in this energy range ( 10 MeV). The experiment showed that the pulse height defect affects the measured spectra the way that the peak energies from the C interactions are shifted by about 0.5 MeV to higher values when the calibration from the Si interactions is used.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
—
OECD FORD obor
10304 - Nuclear physics
Návaznosti výsledku
Projekt
<a href="/cs/project/8X23035" target="_blank" >8X23035: SiC Timepix detector</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2025
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
ANIMMA 2025 – Advancements in Nuclear Instrumentation Measurement Methods and their Applications
ISBN
—
ISSN
2100-014X
e-ISSN
2100-014X
Počet stran výsledku
5
Strana od-do
—
Název nakladatele
EPJ Web of Conferences
Místo vydání
Les Ulis Cedex A
Místo konání akce
Valencia
Datum konání akce
9. 6. 2025
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—