The Czech multicenter dosimetric and treatment planning intercomparison of IMRT with a pelvic phantom
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F86652052%3A_____%2F14%3A%230000225" target="_blank" >RIV/86652052:_____/14:#0000225 - isvavai.cz</a>
Výsledek na webu
<a href="http://www.postersessiononline.eu/pr/aula_poster.asp?congreso=336742579&direccion_posters=Seleccion&pagina_posters=26&ordenacion=autor" target="_blank" >http://www.postersessiononline.eu/pr/aula_poster.asp?congreso=336742579&direccion_posters=Seleccion&pagina_posters=26&ordenacion=autor</a>
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
The Czech multicenter dosimetric and treatment planning intercomparison of IMRT with a pelvic phantom
Popis výsledku v původním jazyce
The aim of the project was to develop a program to assess the performance of prostate IMRT treatment in the Czech Republic. All Czech departments (17) where prostate patients are treated participated. End-to-end test with pelvic phantom including inhomogeneities has been performed. In total 21 IMRT treatment plans (including 4 IMAT plans) were evaluated. Pelvic phantom with contrast structures and defined PTV was developed to represent a standard prostate patient indicated for IMRT treatment. Agreementwith treatment planning system for dosimetric values (beam calibration, dose in PTV and OAR according to the planned/prescribed dose measured with ion chambers, gamma evaluation of planar dose distribution with gafchromic films) and non-dosimetric values(CT numbers to RED conversion, volumes calculation) was evaluated. DVHs were compared for all plans.
Název v anglickém jazyce
The Czech multicenter dosimetric and treatment planning intercomparison of IMRT with a pelvic phantom
Popis výsledku anglicky
The aim of the project was to develop a program to assess the performance of prostate IMRT treatment in the Czech Republic. All Czech departments (17) where prostate patients are treated participated. End-to-end test with pelvic phantom including inhomogeneities has been performed. In total 21 IMRT treatment plans (including 4 IMAT plans) were evaluated. Pelvic phantom with contrast structures and defined PTV was developed to represent a standard prostate patient indicated for IMRT treatment. Agreementwith treatment planning system for dosimetric values (beam calibration, dose in PTV and OAR according to the planned/prescribed dose measured with ion chambers, gamma evaluation of planar dose distribution with gafchromic films) and non-dosimetric values(CT numbers to RED conversion, volumes calculation) was evaluated. DVHs were compared for all plans.
Klasifikace
Druh
O - Ostatní výsledky
CEP obor
FS - Lékařská zařízení, přístroje a vybavení
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/TB01SUJB071" target="_blank" >TB01SUJB071: Výzkum ozáření populace a optimalizace radiační ochrany při lékařském ozáření v České republice</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2014
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů