Laboratory reconstruction of personal doses in interventional cardiology/radiology
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F86652052%3A_____%2F14%3A%230000246" target="_blank" >RIV/86652052:_____/14:#0000246 - isvavai.cz</a>
Výsledek na webu
<a href="http://www.dro2014.sk/files/zbornik_abstraktov.pdf" target="_blank" >http://www.dro2014.sk/files/zbornik_abstraktov.pdf</a>
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Laboratory reconstruction of personal doses in interventional cardiology/radiology
Popis výsledku v původním jazyce
People working in interventional cardiology represent one of the most occupationally exposed groups. By the ICRP 120 recommendation these workers should be equipped with two personal dosimeters, one of them worn on the neck outside the protective collarand the second one on the trunk under the protective apron. It is proved that this double dosimetry method provides more precise estimate of effective dose. A common attitude to estimate eye lens dose with help of standard personal dosimeters worn on thebody usually calibrated in the quantities Hp(0.07) or Hp(10), where it is supposed to wear the dosimeter in the position representative enough to assess the irradiation of the eye lens. Another possibility, how to determine eye lens dose, is to use a special dosimeter worn in the proximity of the eye.
Název v anglickém jazyce
Laboratory reconstruction of personal doses in interventional cardiology/radiology
Popis výsledku anglicky
People working in interventional cardiology represent one of the most occupationally exposed groups. By the ICRP 120 recommendation these workers should be equipped with two personal dosimeters, one of them worn on the neck outside the protective collarand the second one on the trunk under the protective apron. It is proved that this double dosimetry method provides more precise estimate of effective dose. A common attitude to estimate eye lens dose with help of standard personal dosimeters worn on thebody usually calibrated in the quantities Hp(0.07) or Hp(10), where it is supposed to wear the dosimeter in the position representative enough to assess the irradiation of the eye lens. Another possibility, how to determine eye lens dose, is to use a special dosimeter worn in the proximity of the eye.
Klasifikace
Druh
O - Ostatní výsledky
CEP obor
FS - Lékařská zařízení, přístroje a vybavení
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2014
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů