Intercomparison of Gafchromic(TM) films, TL detectors and TL foils for the measurements of skin dose in interventional radiology
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F86652052%3A_____%2F14%3A%230000279" target="_blank" >RIV/86652052:_____/14:#0000279 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1016/j.radmeas.2014.04.008" target="_blank" >http://dx.doi.org/10.1016/j.radmeas.2014.04.008</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.radmeas.2014.04.008" target="_blank" >10.1016/j.radmeas.2014.04.008</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Intercomparison of Gafchromic(TM) films, TL detectors and TL foils for the measurements of skin dose in interventional radiology
Popis výsledku v původním jazyce
Several passive solid state dosemeters, such as Gafchromic(TM) films and thermoluminescence CM detectors, are used to estimate and monitor patient skin doses in interventional radiology. To determine the suitability of XR-TypeR Gafchromic(TM) films and of detectors based on TL materials: pellets, chips and foils to measure skin dose, an intercomparison exercise has been organized within European Dosimetry Radiation Group - Working Group 12 "European Medical ALARA Network" (EURADOS WG12). To test response detectors were exposed to X-ray beams of energies and qualities applied clinically. A blind test was also performed to investigate the accuracy of the dose estimate by detectors exposed to unknown doses. We found the response of films to be strongly dependent on beam quality and filtration (increasing by up to 80% with respect to reference beam quality). The response of TL detectors was found to be less dependent on beam quality (less than 25% variation), with TL. foils showing less th
Název v anglickém jazyce
Intercomparison of Gafchromic(TM) films, TL detectors and TL foils for the measurements of skin dose in interventional radiology
Popis výsledku anglicky
Several passive solid state dosemeters, such as Gafchromic(TM) films and thermoluminescence CM detectors, are used to estimate and monitor patient skin doses in interventional radiology. To determine the suitability of XR-TypeR Gafchromic(TM) films and of detectors based on TL materials: pellets, chips and foils to measure skin dose, an intercomparison exercise has been organized within European Dosimetry Radiation Group - Working Group 12 "European Medical ALARA Network" (EURADOS WG12). To test response detectors were exposed to X-ray beams of energies and qualities applied clinically. A blind test was also performed to investigate the accuracy of the dose estimate by detectors exposed to unknown doses. We found the response of films to be strongly dependent on beam quality and filtration (increasing by up to 80% with respect to reference beam quality). The response of TL detectors was found to be less dependent on beam quality (less than 25% variation), with TL. foils showing less th
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
FS - Lékařská zařízení, přístroje a vybavení
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2014
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Raditation Measurements
ISSN
1350-4487
e-ISSN
—
Svazek periodika
71
Číslo periodika v rámci svazku
SI
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
5
Strana od-do
282-286
Kód UT WoS článku
000347494400061
EID výsledku v databázi Scopus
—