Monte Carlo simulation of the Bremsstrahlung radiation for the measurement ofan internal contamination with pure-beta emitters in vivo
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F86652052%3A_____%2F16%3AN0000022" target="_blank" >RIV/86652052:_____/16:N0000022 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1093/rpd/ncv427" target="_blank" >http://dx.doi.org/10.1093/rpd/ncv427</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1093/rpd/ncv427" target="_blank" >10.1093/rpd/ncv427</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Monte Carlo simulation of the Bremsstrahlung radiation for the measurement ofan internal contamination with pure-beta emitters in vivo
Popis výsledku v původním jazyce
Rapid measurement techniques are required for a large-scale emergency monitoring of people. In vivo measurement of the bremsstrahlung radiation produced by incorporated pure-beta emitters can offer a rapid technique for the determination of such radionuclides in the human body. This work presents a method for the calibration of spectrometers, based on the use of UPh-02T (so-called IGOR) phantom and specific 90Sr/90Y sources, which can account for recent as well as previous contaminations. The process of the whole- and partial-body counter calibration in combination with application of a Monte Carlo code offers readily extension also to other pure-beta emitters and various exposure scenarios.
Název v anglickém jazyce
Monte Carlo simulation of the Bremsstrahlung radiation for the measurement ofan internal contamination with pure-beta emitters in vivo
Popis výsledku anglicky
Rapid measurement techniques are required for a large-scale emergency monitoring of people. In vivo measurement of the bremsstrahlung radiation produced by incorporated pure-beta emitters can offer a rapid technique for the determination of such radionuclides in the human body. This work presents a method for the calibration of spectrometers, based on the use of UPh-02T (so-called IGOR) phantom and specific 90Sr/90Y sources, which can account for recent as well as previous contaminations. The process of the whole- and partial-body counter calibration in combination with application of a Monte Carlo code offers readily extension also to other pure-beta emitters and various exposure scenarios.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
DL - Jaderné odpady, radioaktivní znečištění a kontrola
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/VF20102015014" target="_blank" >VF20102015014: Výzkum pokročilých metod detekce, stanovení a následné zvládnutí radioaktivní kontaminace</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2016
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Radiation Protection Dosimetry
ISSN
0144-8420
e-ISSN
—
Svazek periodika
170
Číslo periodika v rámci svazku
1-4
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
5
Strana od-do
354-358
Kód UT WoS článku
000384713300080