An analytical X-ray CdTe detector response matrix for incomplete charge collection correction for photon energies up to 300 keV
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F86652052%3A_____%2F18%3AN0000039" target="_blank" >RIV/86652052:_____/18:N0000039 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1016/j.radphyschem.2018.01.003" target="_blank" >http://dx.doi.org/10.1016/j.radphyschem.2018.01.003</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.radphyschem.2018.01.003" target="_blank" >10.1016/j.radphyschem.2018.01.003</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
An analytical X-ray CdTe detector response matrix for incomplete charge collection correction for photon energies up to 300 keV
Popis výsledku v původním jazyce
The distortion of the measured energy spectra caused by incomplete charge collection in the CdTe detectors was corrected in a similar way as the other general measuring artefacts, i.e., with the help of an analytically built detector response matrix. The effect of incomplete charge collection was described quantitatively by the Hecht equation and was incorporated into the detector response matrix. The parameters of the Hecht equation, the mean free paths of electrons and holes, were optimized for a particular detector. The performance of the model was tested on a line spectra and an energy resolution of 0.96 % +/- 0.28 % was achieved at 344.27 keV. Application of the correction to measured X-ray spectra leads to quantitative and qualitative agreement of the measured corrected spectra with the theoretically calculated ones. The method was applied and verified for X-ray spectra up to 300 kV.
Název v anglickém jazyce
An analytical X-ray CdTe detector response matrix for incomplete charge collection correction for photon energies up to 300 keV
Popis výsledku anglicky
The distortion of the measured energy spectra caused by incomplete charge collection in the CdTe detectors was corrected in a similar way as the other general measuring artefacts, i.e., with the help of an analytically built detector response matrix. The effect of incomplete charge collection was described quantitatively by the Hecht equation and was incorporated into the detector response matrix. The parameters of the Hecht equation, the mean free paths of electrons and holes, were optimized for a particular detector. The performance of the model was tested on a line spectra and an energy resolution of 0.96 % +/- 0.28 % was achieved at 344.27 keV. Application of the correction to measured X-ray spectra leads to quantitative and qualitative agreement of the measured corrected spectra with the theoretically calculated ones. The method was applied and verified for X-ray spectra up to 300 kV.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
10304 - Nuclear physics
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2018
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Radiation Physics and Chemistry
ISSN
0969-806X
e-ISSN
—
Svazek periodika
146
Číslo periodika v rámci svazku
May
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
8
Strana od-do
26-33
Kód UT WoS článku
000428487600005
EID výsledku v databázi Scopus
2-s2.0-85041457391