Assessing 131I in thyroid by non-spectroscopic instruments - A European intercomparison exercise
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F86652052%3A_____%2F19%3AN0000023" target="_blank" >RIV/86652052:_____/19:N0000023 - isvavai.cz</a>
Výsledek na webu
<a href="https://www.sciencedirect.com/science/article/abs/pii/S1350448718307807" target="_blank" >https://www.sciencedirect.com/science/article/abs/pii/S1350448718307807</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.radmeas.2019.04.018" target="_blank" >10.1016/j.radmeas.2019.04.018</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Assessing 131I in thyroid by non-spectroscopic instruments - A European intercomparison exercise
Popis výsledku v původním jazyce
One of the issues of the Open Project for the European Radiation Research Area (OPERRA) was human thyroid monitoring in case of a large scale nuclear accident. This issue was covered in task 5.4 as project “CaThyMARA” (Child and Adult Thyroid Monitoring After Reactor Accident), which included several aspects of thyroid monitoring, e.g. screening of facilities able to perform thyroid monitoring in the European countries, dose estimation, modelling of detector response, and two intercomparison exercises. The intercomparison described in this paper focused on thyroid monitoring by non-spectrometric instruments, including gamma cameras and other instruments that were considered available for measurements made by members of the public. A total of 12 facilities from 7 European countries have participated and 43 various measuring devices have been evaluated. The main conclusion of this intercomparison is that the ability to make assessments of 131I activity in the thyroid to the exposed population after an accidental release must, on the average, be considered as good among the European laboratories taking part in this study. This intercomparison also gave the participants the possibility to calibrate the measuring devices for thyroid measurements of children where this procedure was not available before. A comprehensive report of the intercomparison is given.
Název v anglickém jazyce
Assessing 131I in thyroid by non-spectroscopic instruments - A European intercomparison exercise
Popis výsledku anglicky
One of the issues of the Open Project for the European Radiation Research Area (OPERRA) was human thyroid monitoring in case of a large scale nuclear accident. This issue was covered in task 5.4 as project “CaThyMARA” (Child and Adult Thyroid Monitoring After Reactor Accident), which included several aspects of thyroid monitoring, e.g. screening of facilities able to perform thyroid monitoring in the European countries, dose estimation, modelling of detector response, and two intercomparison exercises. The intercomparison described in this paper focused on thyroid monitoring by non-spectrometric instruments, including gamma cameras and other instruments that were considered available for measurements made by members of the public. A total of 12 facilities from 7 European countries have participated and 43 various measuring devices have been evaluated. The main conclusion of this intercomparison is that the ability to make assessments of 131I activity in the thyroid to the exposed population after an accidental release must, on the average, be considered as good among the European laboratories taking part in this study. This intercomparison also gave the participants the possibility to calibrate the measuring devices for thyroid measurements of children where this procedure was not available before. A comprehensive report of the intercomparison is given.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
10511 - Environmental sciences (social aspects to be 5.7)
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2019
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Radiation Measurements
ISSN
1350-4487
e-ISSN
—
Svazek periodika
128
Číslo periodika v rámci svazku
September
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
6
Strana od-do
106115
Kód UT WoS článku
000497891200016
EID výsledku v databázi Scopus
2-s2.0-85061425457