Solenoid coil for mouse-model MRI with a clinical 3-Tesla imager: body imaging
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00023001%3A_____%2F09%3A00002090" target="_blank" >RIV/00023001:_____/09:00002090 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Solenoid coil for mouse-model MRI with a clinical 3-Tesla imager: body imaging
Popis výsledku v původním jazyce
A solenoid coil was built for magnetic resonance imaging of the mice. A coil prototype composed of 5 turns, with a length of 4 cm and 2.5 cm radius was developed to acquire (whole) body mouse magnetic resonance images at 130 MHz and an insertable gradient coil set. Coil performance was measured using the Q factor for both the loaded and unloaded cases were 161.67 and 178.03, respectively. These Q factors compare very well with those values reported in the literature. The images were acquired with a clinical 3 T system equipped with a custom-built gradient insert coil and gradient echo image sequence. Both phantom and in vivo images showed good signal-to-noise ratio and uniformity. The electromagnetic interaction between the insertable coils and the solenoidal coil is poor, and no image artefacts are present in the whole-body image of the mouse. This preliminary experience has shown that consistent high quality MR images of the mice can be obtained using this particular hardware configu
Název v anglickém jazyce
Solenoid coil for mouse-model MRI with a clinical 3-Tesla imager: body imaging
Popis výsledku anglicky
A solenoid coil was built for magnetic resonance imaging of the mice. A coil prototype composed of 5 turns, with a length of 4 cm and 2.5 cm radius was developed to acquire (whole) body mouse magnetic resonance images at 130 MHz and an insertable gradient coil set. Coil performance was measured using the Q factor for both the loaded and unloaded cases were 161.67 and 178.03, respectively. These Q factors compare very well with those values reported in the literature. The images were acquired with a clinical 3 T system equipped with a custom-built gradient insert coil and gradient echo image sequence. Both phantom and in vivo images showed good signal-to-noise ratio and uniformity. The electromagnetic interaction between the insertable coils and the solenoidal coil is poor, and no image artefacts are present in the whole-body image of the mouse. This preliminary experience has shown that consistent high quality MR images of the mice can be obtained using this particular hardware configu
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
FS - Lékařská zařízení, přístroje a vybavení
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
N - Vyzkumna aktivita podporovana z neverejnych zdroju
Ostatní
Rok uplatnění
2009
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Revista mexicana de física
ISSN
0035-001X
e-ISSN
—
Svazek periodika
55
Číslo periodika v rámci svazku
2
Stát vydavatele periodika
MX - Spojené státy mexické
Počet stran výsledku
5
Strana od-do
—
Kód UT WoS článku
000265620100011
EID výsledku v databázi Scopus
—