Gold vs. platinum-iridium tip catheter for cavotricuspid isthmus ablation: the AURUM 8 study
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00023001%3A_____%2F11%3A00002507" target="_blank" >RIV/00023001:_____/11:00002507 - isvavai.cz</a>
Výsledek na webu
<a href="http://europace.oxfordjournals.org/content/13/1/102" target="_blank" >http://europace.oxfordjournals.org/content/13/1/102</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1093/europace/euq339" target="_blank" >10.1093/europace/euq339</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Gold vs. platinum-iridium tip catheter for cavotricuspid isthmus ablation: the AURUM 8 study
Popis výsledku v původním jazyce
Aims Gold electrodes have the theoretical advantage of creating bigger lesions than platinum-iridium (Pt-Ir) electrodes. We performed a prospective randomized study to compare the clinical efficacy of standard 8 mm Pt-Ir tip catheter (control) and 8 mm gold-tip catheters in the ablation of the cavotricuspid isthmus (CTI)-dependent atrial flutter. Methods and results A total of 463 patients undergoing CTI ablation in 19 clinical centres were randomized to receive the treatment by gold-tip or control catheter. The primary endpoint was cumulative radiofrequency (RF) application duration until achieving bidirectional CTI block. It did not differ significantly for the two catheters. The gold-tip catheter was, however, associated with a higher ablation success rate (94.3 vs. 89.0%, P = 0.042) and a substantially lower incidence of char and coagulum formation (4.8 vs. 37.9%, P < 0.001), which required exchange of 1 gold-tip (0.4%) and 10 control catheters (4.6%, P = 0.005). The gold-tip cathe
Název v anglickém jazyce
Gold vs. platinum-iridium tip catheter for cavotricuspid isthmus ablation: the AURUM 8 study
Popis výsledku anglicky
Aims Gold electrodes have the theoretical advantage of creating bigger lesions than platinum-iridium (Pt-Ir) electrodes. We performed a prospective randomized study to compare the clinical efficacy of standard 8 mm Pt-Ir tip catheter (control) and 8 mm gold-tip catheters in the ablation of the cavotricuspid isthmus (CTI)-dependent atrial flutter. Methods and results A total of 463 patients undergoing CTI ablation in 19 clinical centres were randomized to receive the treatment by gold-tip or control catheter. The primary endpoint was cumulative radiofrequency (RF) application duration until achieving bidirectional CTI block. It did not differ significantly for the two catheters. The gold-tip catheter was, however, associated with a higher ablation success rate (94.3 vs. 89.0%, P = 0.042) and a substantially lower incidence of char and coagulum formation (4.8 vs. 37.9%, P < 0.001), which required exchange of 1 gold-tip (0.4%) and 10 control catheters (4.6%, P = 0.005). The gold-tip cathe
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
FA - Kardiovaskulární nemoci včetně kardiochirurgie
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
N - Vyzkumna aktivita podporovana z neverejnych zdroju
Ostatní
Rok uplatnění
2011
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Europace
ISSN
1099-5129
e-ISSN
—
Svazek periodika
13
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
7
Strana od-do
102-108
Kód UT WoS článku
000285417000021
EID výsledku v databázi Scopus
—