Retrieval of Long-Term Implanted Leadless Pacemakers: A Single-Center Experience
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00023884%3A_____%2F20%3A00008739" target="_blank" >RIV/00023884:_____/20:00008739 - isvavai.cz</a>
Výsledek na webu
<a href="https://www.sciencedirect.com/science/article/abs/pii/S2405500X20305065" target="_blank" >https://www.sciencedirect.com/science/article/abs/pii/S2405500X20305065</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.jacep.2020.06.010" target="_blank" >10.1016/j.jacep.2020.06.010</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Retrieval of Long-Term Implanted Leadless Pacemakers: A Single-Center Experience
Popis výsledku v původním jazyce
OBJECTIVES The authors report their single-center experience with the retrieval and replacement of long-term implanted leadless cardiac pacemakers (LPs) and the factors contributing to success. BACKGROUND LPs are a clinically effective and safe alternative to standard transvenous pacemakers for single-chamber ventricular pacing. However, the feasibility of retrieving and replacing long-term implanted LPs is not well known. METHODS A total of 34 patients with implanted Nanostim LPs (mean implantation duration 1,570 +/- 479 days) subsequently underwent retrieval. On the basis of fluoroscopic imaging, the cohort was divided into 2 groups: those with remarkable swinging movement (SM) of the LP docking button (n = 25) and those without SM (n = 9). RESULTS The overall LP retrieval success rate was 85% (n = 29). New leadless devices were implanted in 27 patients immediately after LP retrieval. No procedure-related adverse events occurred. The rate of successful LP retrieval was significantly higher in the SM group (25 of 25 [100%]) than in the no-SM group (4 of 9 [44%]) (p < 0.001), and fluoroscopy time during the retrieval procedure was shorter in the SM group (12.7 +/- 8.6 min) than in the no-SM group (45.6 +/- 19 min) (p < 0.001). CONCLUSIONS This study demonstrated the feasibility and safety of retrieval of long-term implanted LPs after a mean duration of 4 years. Retrieval was most successful in patients whose docking buttons exhibited significant SM. (C) 2020 by the American College of Cardiology Foundation.
Název v anglickém jazyce
Retrieval of Long-Term Implanted Leadless Pacemakers: A Single-Center Experience
Popis výsledku anglicky
OBJECTIVES The authors report their single-center experience with the retrieval and replacement of long-term implanted leadless cardiac pacemakers (LPs) and the factors contributing to success. BACKGROUND LPs are a clinically effective and safe alternative to standard transvenous pacemakers for single-chamber ventricular pacing. However, the feasibility of retrieving and replacing long-term implanted LPs is not well known. METHODS A total of 34 patients with implanted Nanostim LPs (mean implantation duration 1,570 +/- 479 days) subsequently underwent retrieval. On the basis of fluoroscopic imaging, the cohort was divided into 2 groups: those with remarkable swinging movement (SM) of the LP docking button (n = 25) and those without SM (n = 9). RESULTS The overall LP retrieval success rate was 85% (n = 29). New leadless devices were implanted in 27 patients immediately after LP retrieval. No procedure-related adverse events occurred. The rate of successful LP retrieval was significantly higher in the SM group (25 of 25 [100%]) than in the no-SM group (4 of 9 [44%]) (p < 0.001), and fluoroscopy time during the retrieval procedure was shorter in the SM group (12.7 +/- 8.6 min) than in the no-SM group (45.6 +/- 19 min) (p < 0.001). CONCLUSIONS This study demonstrated the feasibility and safety of retrieval of long-term implanted LPs after a mean duration of 4 years. Retrieval was most successful in patients whose docking buttons exhibited significant SM. (C) 2020 by the American College of Cardiology Foundation.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
30201 - Cardiac and Cardiovascular systems
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2020
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
JACC: Clinical Electrophysiology
ISSN
2405-500X
e-ISSN
—
Svazek periodika
6
Číslo periodika v rámci svazku
14
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
8
Strana od-do
1744-1751
Kód UT WoS článku
000602753200002
EID výsledku v databázi Scopus
2-s2.0-85090482155