Final Report on EUROMET.T-K3: Regional key comparison of the realisations of the ITS-90 from 83.8058 K to 692.677 K
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F07%3A%230000449" target="_blank" >RIV/00177016:_____/07:#0000449 - isvavai.cz</a>
Výsledek na webu
<a href="http://iopscience.iop.org/0026-1394/44/1A/03001?fromSearchPage=true" target="_blank" >http://iopscience.iop.org/0026-1394/44/1A/03001?fromSearchPage=true</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1088/0026-1394/44/1A/03001" target="_blank" >10.1088/0026-1394/44/1A/03001</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Final Report on EUROMET.T-K3: Regional key comparison of the realisations of the ITS-90 from 83.8058 K to 692.677 K
Popis výsledku v původním jazyce
This report presents the results of the EUROMET regional key comparison corresponding to CCT-K3 and covering the temperature range from 83.8058 K (triple point of Ar) to 692.77 K (freezing point of Zn). The comparison involved the six NMIs previously involved in CCT-K3 (BNM-INM/CNAM, SMU, IMGC, NMi-VSL, NPL, PTB) and 18 European national laboratories. The comparison was divided into five different loops coordinated by a co-pilot chosen among the laboratories having participated in the CCT-K3 comparison.BNM-INM/CNAM played the role of pilot in establishing the link between the five loops. In each loop, an artefact in the form of a standard platinum resistance thermometer (SPRT) was circulated among the participant laboratories. In order to have sufficient information about the possible drift of the SPRTs, the co-pilots have performed a calibration over the full temperature range at the beginning and at the end of the loop.
Název v anglickém jazyce
Final Report on EUROMET.T-K3: Regional key comparison of the realisations of the ITS-90 from 83.8058 K to 692.677 K
Popis výsledku anglicky
This report presents the results of the EUROMET regional key comparison corresponding to CCT-K3 and covering the temperature range from 83.8058 K (triple point of Ar) to 692.77 K (freezing point of Zn). The comparison involved the six NMIs previously involved in CCT-K3 (BNM-INM/CNAM, SMU, IMGC, NMi-VSL, NPL, PTB) and 18 European national laboratories. The comparison was divided into five different loops coordinated by a co-pilot chosen among the laboratories having participated in the CCT-K3 comparison.BNM-INM/CNAM played the role of pilot in establishing the link between the five loops. In each loop, an artefact in the form of a standard platinum resistance thermometer (SPRT) was circulated among the participant laboratories. In order to have sufficient information about the possible drift of the SPRTs, the co-pilots have performed a calibration over the full temperature range at the beginning and at the end of the loop.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
BJ - Termodynamika
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
V - Vyzkumna aktivita podporovana z jinych verejnych zdroju
Ostatní
Rok uplatnění
2007
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Metrologia
ISSN
0026-1394
e-ISSN
—
Svazek periodika
2007
Číslo periodika v rámci svazku
44
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
12
Strana od-do
3056-68
Kód UT WoS článku
—