The laboratory setup for calibrating of the surface temperature sensors
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F10%3A%230000478" target="_blank" >RIV/00177016:_____/10:#0000478 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
The laboratory setup for calibrating of the surface temperature sensors
Popis výsledku v původním jazyce
The calibration of the surface temperature sensors depends on number of influencing parameters including the calibrated sensor, design of the surface, ambient environment and methods of the calibration. The results also often depend on the operator experiences. In this article will be presented the laboratory setup for surface temperature calibration at ČMI. The design is based on the block of compact material with 4pc of industrial platinum resistance thermometers (IPRT) to measure the temperature gradient along the block. These measurements were used for extrapolation of the surface temperature. In the experimental part will be showed the comparison of this block with the commercial system used at ČMI. It will be also shown the influence of differentblock materials and different designs to the results. The investigation on the external environment was made. The results were compared with numerical analysis and the uncertainty budged will be presented.
Název v anglickém jazyce
The laboratory setup for calibrating of the surface temperature sensors
Popis výsledku anglicky
The calibration of the surface temperature sensors depends on number of influencing parameters including the calibrated sensor, design of the surface, ambient environment and methods of the calibration. The results also often depend on the operator experiences. In this article will be presented the laboratory setup for surface temperature calibration at ČMI. The design is based on the block of compact material with 4pc of industrial platinum resistance thermometers (IPRT) to measure the temperature gradient along the block. These measurements were used for extrapolation of the surface temperature. In the experimental part will be showed the comparison of this block with the commercial system used at ČMI. It will be also shown the influence of differentblock materials and different designs to the results. The investigation on the external environment was made. The results were compared with numerical analysis and the uncertainty budged will be presented.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
BJ - Termodynamika
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
V - Vyzkumna aktivita podporovana z jinych verejnych zdroju
Ostatní
Rok uplatnění
2010
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Tempmeko
ISBN
978-961-243-146-4
ISSN
—
e-ISSN
—
Počet stran výsledku
1
Strana od-do
251-251
Název nakladatele
IMEKO TC 12
Místo vydání
Portorož
Místo konání akce
Portorož
Datum konání akce
1. 1. 2010
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—