Comparison at the sub-100 fW optical power level of calibrating a single-photon detector using a high-sensitive, low-noise silicon photodiode and the double attenuator technique
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F16%3AN0000002" target="_blank" >RIV/00177016:_____/16:N0000002 - isvavai.cz</a>
Výsledek na webu
<a href="http://iopscience.iop.org/article/10.1088/0026-1394/53/4/1115/pdf" target="_blank" >http://iopscience.iop.org/article/10.1088/0026-1394/53/4/1115/pdf</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1088/0026-1394/53/4/1115" target="_blank" >10.1088/0026-1394/53/4/1115</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Comparison at the sub-100 fW optical power level of calibrating a single-photon detector using a high-sensitive, low-noise silicon photodiode and the double attenuator technique
Popis výsledku v původním jazyce
A comparison down to sub-100-fW optical power level was carried out between a low-noise Silicon photodiode and a low optical flux measurement facility based on a double attenuator technique. The comparison was carried out via a silicon single-photon avalanche diode (Si-SPAD), which acted as transfer standard. The measurements were performed at a wavelength of 770 nm using an attenuated laser as a radiation source at optical power levels between approximately 86 fW and approximately 1325 fW, corresponding to approximately 330 000 photons s−1 and approximately 5.2 × 106 photons s−1, respectively. The mean relative deviation of the detection efficiencies of the Si-SPAD, determined by the Si-photodiode and the low optical flux measurement facility, i.e. between two completely independent traceability routes, was < 0.2%, thus well within the combined standard uncertainty of the two measurements. To our knowledge, this is the first comparison for the detection efficiency of a single photon detector using a direct optical flux measurement by a conventional Si-photodiode at such low power levels.
Název v anglickém jazyce
Comparison at the sub-100 fW optical power level of calibrating a single-photon detector using a high-sensitive, low-noise silicon photodiode and the double attenuator technique
Popis výsledku anglicky
A comparison down to sub-100-fW optical power level was carried out between a low-noise Silicon photodiode and a low optical flux measurement facility based on a double attenuator technique. The comparison was carried out via a silicon single-photon avalanche diode (Si-SPAD), which acted as transfer standard. The measurements were performed at a wavelength of 770 nm using an attenuated laser as a radiation source at optical power levels between approximately 86 fW and approximately 1325 fW, corresponding to approximately 330 000 photons s−1 and approximately 5.2 × 106 photons s−1, respectively. The mean relative deviation of the detection efficiencies of the Si-SPAD, determined by the Si-photodiode and the low optical flux measurement facility, i.e. between two completely independent traceability routes, was < 0.2%, thus well within the combined standard uncertainty of the two measurements. To our knowledge, this is the first comparison for the detection efficiency of a single photon detector using a direct optical flux measurement by a conventional Si-photodiode at such low power levels.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
10306 - Optics (including laser optics and quantum optics)
Návaznosti výsledku
Projekt
<a href="/cs/project/7AX13015" target="_blank" >7AX13015: Single-photon sources for quantum technologies</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2016
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Metrologia
ISSN
1681-7575
e-ISSN
1681-7575
Svazek periodika
53
Číslo periodika v rámci svazku
4
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
7
Strana od-do
1115-1122
Kód UT WoS článku
000381914600014
EID výsledku v databázi Scopus
2-s2.0-84983637803