EURAMET.M.P-S9 / EURAMET 1170, Loop 2. Comparison in the negative gauge pressure range −950 to 0 hPa
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F17%3AN0000066" target="_blank" >RIV/00177016:_____/17:N0000066 - isvavai.cz</a>
Výsledek na webu
<a href="https://www.bipm.org/utils/common/pdf/final_reports/M/P-S9/EURAMET.M.P-S9_Loop_2.pdf" target="_blank" >https://www.bipm.org/utils/common/pdf/final_reports/M/P-S9/EURAMET.M.P-S9_Loop_2.pdf</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1088/0026-1394/54/1A/07004" target="_blank" >10.1088/0026-1394/54/1A/07004</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
EURAMET.M.P-S9 / EURAMET 1170, Loop 2. Comparison in the negative gauge pressure range −950 to 0 hPa
Popis výsledku v původním jazyce
This report gives the results of a comparison of pressure standards of seven European National Metrology institutes in the range of negative gauge pressure from -950 hPa to 0 hPa, registered as a supplementary comparison EURAMET.M.P-S9. This comparison was carried out from January 2011 to March 2012. The transfer standard used was a pressure monitor RPM4 A160Ks, with a resolution of 0.1 Pa. The reference values have been determined from the weighted mean of the deviations reported by the participants for each specified pressure. Seventy three of the seventy seven values reported by the laboratories agree with the reference values within the expanded uncertainties.
Název v anglickém jazyce
EURAMET.M.P-S9 / EURAMET 1170, Loop 2. Comparison in the negative gauge pressure range −950 to 0 hPa
Popis výsledku anglicky
This report gives the results of a comparison of pressure standards of seven European National Metrology institutes in the range of negative gauge pressure from -950 hPa to 0 hPa, registered as a supplementary comparison EURAMET.M.P-S9. This comparison was carried out from January 2011 to March 2012. The transfer standard used was a pressure monitor RPM4 A160Ks, with a resolution of 0.1 Pa. The reference values have been determined from the weighted mean of the deviations reported by the participants for each specified pressure. Seventy three of the seventy seven values reported by the laboratories agree with the reference values within the expanded uncertainties.
Klasifikace
Druh
J<sub>ost</sub> - Ostatní články v recenzovaných periodicích
CEP obor
—
OECD FORD obor
20301 - Mechanical engineering
Návaznosti výsledku
Projekt
—
Návaznosti
V - Vyzkumna aktivita podporovana z jinych verejnych zdroju
Ostatní
Rok uplatnění
2017
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Metrologia
ISSN
0026-1394
e-ISSN
1681-7575
Svazek periodika
54
Číslo periodika v rámci svazku
Tech. Suppl. 07004
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
21
Strana od-do
1-21