X-ray spectra emitted by the electronic brachytherapy unit Papillon 50 measured with a CdTe spectrometer
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F25%3AN0000109" target="_blank" >RIV/00177016:_____/25:N0000109 - isvavai.cz</a>
Výsledek na webu
<a href="https://doi.org/10.5114/jcb.2025.154104" target="_blank" >https://doi.org/10.5114/jcb.2025.154104</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.5114/jcb.2025.154104" target="_blank" >10.5114/jcb.2025.154104</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
X-ray spectra emitted by the electronic brachytherapy unit Papillon 50 measured with a CdTe spectrometer
Popis výsledku v původním jazyce
Purpose: The aim of this paper was to present the measured photon fluence spectra emitted by the electronic brachytherapy (eBT) unit Papillon 50. Material and methods: Spectra were collected with a compact CdTe spectrometer in air at a distance of 52 cm from the unit output window, for three different applicators attached to the unit and at several angles with respect to the beam axis. Results: The fluence spectra recorded for this device showed lower energy of the photon spectrum end-point than anticipated. The end-point energy was found to be (47.6 ±0.2) keV for the nominal high voltage bias of 50 kV set by the operator. Full-energy peaks in the spectra identified characteristic photons emitted by the rhenium anode, and by molybdenum and silver supposedly present in the inner coating of the applicators. An asymmetry in the half-value layer values of measured spectra was observed. Conclusions: The presented work provides a comprehensive dataset of photon fluence spectra for traceable dosimetry using the Papillon 50 eBT device, which is required, for example, for correcting the energy dependence of detectors or as an input for Monte Carlo calculations. Spectrometry and dose characterization studies are needed for eBT systems.
Název v anglickém jazyce
X-ray spectra emitted by the electronic brachytherapy unit Papillon 50 measured with a CdTe spectrometer
Popis výsledku anglicky
Purpose: The aim of this paper was to present the measured photon fluence spectra emitted by the electronic brachytherapy (eBT) unit Papillon 50. Material and methods: Spectra were collected with a compact CdTe spectrometer in air at a distance of 52 cm from the unit output window, for three different applicators attached to the unit and at several angles with respect to the beam axis. Results: The fluence spectra recorded for this device showed lower energy of the photon spectrum end-point than anticipated. The end-point energy was found to be (47.6 ±0.2) keV for the nominal high voltage bias of 50 kV set by the operator. Full-energy peaks in the spectra identified characteristic photons emitted by the rhenium anode, and by molybdenum and silver supposedly present in the inner coating of the applicators. An asymmetry in the half-value layer values of measured spectra was observed. Conclusions: The presented work provides a comprehensive dataset of photon fluence spectra for traceable dosimetry using the Papillon 50 eBT device, which is required, for example, for correcting the energy dependence of detectors or as an input for Monte Carlo calculations. Spectrometry and dose characterization studies are needed for eBT systems.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
10301 - Atomic, molecular and chemical physics (physics of atoms and molecules including collision, interaction with radiation, magnetic resonances, Mössbauer effect)
Návaznosti výsledku
Projekt
<a href="/cs/project/8B19003" target="_blank" >8B19003: Primary standards and traceable measurement methods for X-ray emitting electronic brachytherapy devices</a><br>
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2025
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Journal of Contemporary Brachytherapy
ISSN
1689-832X
e-ISSN
2081-2841
Svazek periodika
17
Číslo periodika v rámci svazku
4
Stát vydavatele periodika
PL - Polská republika
Počet stran výsledku
8
Strana od-do
—
Kód UT WoS článku
001584231200004
EID výsledku v databázi Scopus
—