Very fast electrophoretic determination of creatinine and uric acid in human urine using a combination of two capillaries with different internal diameters
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216208%3A11120%2F14%3A43907632" target="_blank" >RIV/00216208:11120/14:43907632 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1002/elps.201300293" target="_blank" >http://dx.doi.org/10.1002/elps.201300293</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1002/elps.201300293" target="_blank" >10.1002/elps.201300293</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Very fast electrophoretic determination of creatinine and uric acid in human urine using a combination of two capillaries with different internal diameters
Popis výsledku v původním jazyce
A capillary system formed by combining 25 ?m and 100 ?m id capillaries was used in the short-end injection mode to determine creatinine and uric acid in human urine. The separation was performed at an electric field intensity of 2.3 kV/cm. Creatinine wasdetermined in a background electrolyte with a composition of 20 mM citric acid/NaOH (pH 3.0) and uric acid was determined in 20 mM MES/NaOH (pH 6.0). Under these conditions, migration times of 12.2 s for creatinine and 8.6 s for uric acid were achieved.The LOD value is 2.4 mg/L for creatinine and 0.9 mg/L for uric acid; the RSD for the migration time varies in the range 0.7-1.1% (intra day) to 1.0-7.5% (inter day); RSDs for the peak areas equalled 3.4-4.0% (intra day) and 4.3-4.7% (inter day). The determined creatinine values in 7 urine samples vary in the range 221-1394 mg/L for creatinine and 87-615 mg/L for uric acid. T-test did not reveal any statistically significant difference between the developed CE methodologies and reference
Název v anglickém jazyce
Very fast electrophoretic determination of creatinine and uric acid in human urine using a combination of two capillaries with different internal diameters
Popis výsledku anglicky
A capillary system formed by combining 25 ?m and 100 ?m id capillaries was used in the short-end injection mode to determine creatinine and uric acid in human urine. The separation was performed at an electric field intensity of 2.3 kV/cm. Creatinine wasdetermined in a background electrolyte with a composition of 20 mM citric acid/NaOH (pH 3.0) and uric acid was determined in 20 mM MES/NaOH (pH 6.0). Under these conditions, migration times of 12.2 s for creatinine and 8.6 s for uric acid were achieved.The LOD value is 2.4 mg/L for creatinine and 0.9 mg/L for uric acid; the RSD for the migration time varies in the range 0.7-1.1% (intra day) to 1.0-7.5% (inter day); RSDs for the peak areas equalled 3.4-4.0% (intra day) and 4.3-4.7% (inter day). The determined creatinine values in 7 urine samples vary in the range 221-1394 mg/L for creatinine and 87-615 mg/L for uric acid. T-test did not reveal any statistically significant difference between the developed CE methodologies and reference
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CG - Elektrochemie
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GAP206%2F11%2F0707" target="_blank" >GAP206/11/0707: Voltametrické studium nových iontových kapalin pro jejich využití v kapilární elektroforéze</a><br>
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2014
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Electrophoresis
ISSN
0173-0835
e-ISSN
—
Svazek periodika
35
Číslo periodika v rámci svazku
7
Stát vydavatele periodika
DE - Spolková republika Německo
Počet stran výsledku
6
Strana od-do
956-961
Kód UT WoS článku
000333644600004