Interaction of heparin and tetraarginine in capillary electrophoresis: Implication for analytical applications
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216208%3A11310%2F20%3A10419218" target="_blank" >RIV/00216208:11310/20:10419218 - isvavai.cz</a>
Výsledek na webu
<a href="https://verso.is.cuni.cz/pub/verso.fpl?fname=obd_publikace_handle&handle=YIu1DWJqD0" target="_blank" >https://verso.is.cuni.cz/pub/verso.fpl?fname=obd_publikace_handle&handle=YIu1DWJqD0</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1002/elps.202000011" target="_blank" >10.1002/elps.202000011</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Interaction of heparin and tetraarginine in capillary electrophoresis: Implication for analytical applications
Popis výsledku v původním jazyce
Interactions between heparin and tetraarginine in an acidic background electrolyte were investigated in capillary electrophoresis. The results showed that tetraarginine and heparin form a stable complex that migrates toward the anode immediately after coming into contact. When a zone of tetraarginine at a mg/mL concentration level passes through a zone of heparin at a mu g/mL concentration level, tetraarginine is gradually removed by the formation of the complex that migrates in the opposite direction, thereby decreasing the tetraarginine peak area. The variation of the tetraarginine peak area as a function of the unfractionated heparin concentration was linear within the range 2-20 mu g/mL, which enables us to detect and determine heparin concentrations undetectable with a UV detector. The same behavior was confirmed for low molecular weight heparin.
Název v anglickém jazyce
Interaction of heparin and tetraarginine in capillary electrophoresis: Implication for analytical applications
Popis výsledku anglicky
Interactions between heparin and tetraarginine in an acidic background electrolyte were investigated in capillary electrophoresis. The results showed that tetraarginine and heparin form a stable complex that migrates toward the anode immediately after coming into contact. When a zone of tetraarginine at a mg/mL concentration level passes through a zone of heparin at a mu g/mL concentration level, tetraarginine is gradually removed by the formation of the complex that migrates in the opposite direction, thereby decreasing the tetraarginine peak area. The variation of the tetraarginine peak area as a function of the unfractionated heparin concentration was linear within the range 2-20 mu g/mL, which enables us to detect and determine heparin concentrations undetectable with a UV detector. The same behavior was confirmed for low molecular weight heparin.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
10406 - Analytical chemistry
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach<br>I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2020
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Electrophoresis
ISSN
0173-0835
e-ISSN
—
Svazek periodika
41
Číslo periodika v rámci svazku
21-22
Stát vydavatele periodika
DE - Spolková republika Německo
Počet stran výsledku
6
Strana od-do
1826-1831
Kód UT WoS článku
000556918100001
EID výsledku v databázi Scopus
2-s2.0-85083842198