Influence of the pulsed amperometric detection to the passivation of the working electrode
Popis výsledku v původním jazyce
The degree of passivation of the working electrode during the amperometric detection in flow can be strongly diminished by the application of pulsed amperometry, even without the application of specific cleaning pulses. Three different potential programs were tested and compared to the traditional amperometry with continuously applied potential. All three programs suppressed the passivation in different degree, although the observed efficiency depends also on the conditions of the determination, particularly the pH of the electrolyte. Simple switching between the detection potential and zero potential provided the best results; more complex potential programs shower either the lower efficiency or the low measurement frequency. The detection and quantification limit, obtained using this method, reached 5.8 x 10(-6) mol dm(-3) and 1.9 x 10(-5) mol dm(-3), respectively; their relatively high value is caused by the increase of the background current, inevitably caused by the potential changes.
Název v anglickém jazyce
Influence of the pulsed amperometric detection to the passivation of the working electrode
Popis výsledku anglicky
The degree of passivation of the working electrode during the amperometric detection in flow can be strongly diminished by the application of pulsed amperometry, even without the application of specific cleaning pulses. Three different potential programs were tested and compared to the traditional amperometry with continuously applied potential. All three programs suppressed the passivation in different degree, although the observed efficiency depends also on the conditions of the determination, particularly the pH of the electrolyte. Simple switching between the detection potential and zero potential provided the best results; more complex potential programs shower either the lower efficiency or the low measurement frequency. The detection and quantification limit, obtained using this method, reached 5.8 x 10(-6) mol dm(-3) and 1.9 x 10(-5) mol dm(-3), respectively; their relatively high value is caused by the increase of the background current, inevitably caused by the potential changes.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
10406 - Analytical chemistry
Návaznosti výsledku
Projekt
<a href="/cs/project/GA20-03187S" target="_blank" >GA20-03187S: Studium elektrochemických a adsorpčních dějů na nanostrukturovaných sp2 a sp3 uhlíkových površích ve vztahu k jejich fyzikálně chemickým vlastnostem</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2023
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Monatshefte für Chemie - Chemical Monthly
ISSN
0026-9247
e-ISSN
1434-4475
Svazek periodika
154
Číslo periodika v rámci svazku
9
Stát vydavatele periodika
AT - Rakouská republika
Počet stran výsledku
5
Strana od-do
1019-1023
Kód UT WoS článku
001019483400003
EID výsledku v databázi Scopus
2-s2.0-85163396740