Data processing correction of the irising effect of a fast-gating intensified charge-coupled device on laser-pulse-excited luminescence spectra
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216208%3A11320%2F10%3A10070224" target="_blank" >RIV/00216208:11320/10:10070224 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Data processing correction of the irising effect of a fast-gating intensified charge-coupled device on laser-pulse-excited luminescence spectra
Popis výsledku v původním jazyce
Intensified charge-coupled devices (ICCDs) comprise the advantages of both fast gating detectors and spectrally broad CCDs into one device that enables temporally and spectrally resolved measurements with a few nanosecond resolution. Gating of the measured signal occurs in the image intensifier tube, where a high voltage is applied between the detector photocathode and a microchannel plate electron multiplier. An issue arises in time-resolved luminescence spectroscopy when signal onset characterizationis required. In this case, the transient gate closing process that causes the detected signal always arises in the middle of the ICCD chip regardless of the spectral detection window-the so-called irising effect. We demonstrate that in case when the detection gate width is comparable to the opening/closing time and the gate is pretriggered with respect to the signal onset, the irising effect causes the obtained data to be strongly distorted.
Název v anglickém jazyce
Data processing correction of the irising effect of a fast-gating intensified charge-coupled device on laser-pulse-excited luminescence spectra
Popis výsledku anglicky
Intensified charge-coupled devices (ICCDs) comprise the advantages of both fast gating detectors and spectrally broad CCDs into one device that enables temporally and spectrally resolved measurements with a few nanosecond resolution. Gating of the measured signal occurs in the image intensifier tube, where a high voltage is applied between the detector photocathode and a microchannel plate electron multiplier. An issue arises in time-resolved luminescence spectroscopy when signal onset characterizationis required. In this case, the transient gate closing process that causes the detected signal always arises in the middle of the ICCD chip regardless of the spectral detection window-the so-called irising effect. We demonstrate that in case when the detection gate width is comparable to the opening/closing time and the gate is pretriggered with respect to the signal onset, the irising effect causes the obtained data to be strongly distorted.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
BH - Optika, masery a lasery
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>Z - Vyzkumny zamer (s odkazem do CEZ)<br>S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2010
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Review of Scientific Instruments
ISSN
0034-6748
e-ISSN
—
Svazek periodika
81
Číslo periodika v rámci svazku
6
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
5
Strana od-do
—
Kód UT WoS článku
000280102200004
EID výsledku v databázi Scopus
—