New Virtual Frisch-Grid CdZnTe Detector Design With Sub-Millimeter Spatial Resolution
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216208%3A11320%2F14%3A10283179" target="_blank" >RIV/00216208:11320/14:10283179 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1109/TNS.2014.2348572" target="_blank" >http://dx.doi.org/10.1109/TNS.2014.2348572</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1109/TNS.2014.2348572" target="_blank" >10.1109/TNS.2014.2348572</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
New Virtual Frisch-Grid CdZnTe Detector Design With Sub-Millimeter Spatial Resolution
Popis výsledku v původním jazyce
We evaluated the performance of a position-sensitive virtual Frisch-grid (VFG) CdZnTe detector, mm mm mm, via sensing strips on its side surfaces. Once the signals were collected from the anode, and from four or eight strips attached to the detector's sides, we assessed the anode's energy spectra and derived histograms from the side electrodes to evaluate the feasibility of achieving sub-millimeter spatial resolution in the X-Y plane. Using a highly collimated 30-keV X-ray beam at the National Synchrotron Light Source, and applying corrections to the raw data, we determined the photon-interaction points by conventional Anger logic and via a more sophisticated statistics-based positioning (SBP) algorithm. With the VFG detector's current configuration, we achieved a resolution below 1 mm, even for low-energy X-rays.
Název v anglickém jazyce
New Virtual Frisch-Grid CdZnTe Detector Design With Sub-Millimeter Spatial Resolution
Popis výsledku anglicky
We evaluated the performance of a position-sensitive virtual Frisch-grid (VFG) CdZnTe detector, mm mm mm, via sensing strips on its side surfaces. Once the signals were collected from the anode, and from four or eight strips attached to the detector's sides, we assessed the anode's energy spectra and derived histograms from the side electrodes to evaluate the feasibility of achieving sub-millimeter spatial resolution in the X-Y plane. Using a highly collimated 30-keV X-ray beam at the National Synchrotron Light Source, and applying corrections to the raw data, we determined the photon-interaction points by conventional Anger logic and via a more sophisticated statistics-based positioning (SBP) algorithm. With the VFG detector's current configuration, we achieved a resolution below 1 mm, even for low-energy X-rays.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
BM - Fyzika pevných látek a magnetismus
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2014
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
IEEE Transactions on Nuclear Science
ISSN
0018-9499
e-ISSN
—
Svazek periodika
61
Číslo periodika v rámci svazku
5
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
6
Strana od-do
2567-2572
Kód UT WoS článku
000343980900008
EID výsledku v databázi Scopus
—