Calibration of High-resolution Spectrometer for C III Triplet Measurements
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216208%3A11320%2F15%3A10319832" target="_blank" >RIV/00216208:11320/15:10319832 - isvavai.cz</a>
Výsledek na webu
<a href="http://www.mff.cuni.cz/veda/konference/wds/proc/pdf15/WDS15_47_f2_Tomes.pdf" target="_blank" >http://www.mff.cuni.cz/veda/konference/wds/proc/pdf15/WDS15_47_f2_Tomes.pdf</a>
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Calibration of High-resolution Spectrometer for C III Triplet Measurements
Popis výsledku v původním jazyce
Edge poloidal plasma rotation and ion temperature at the COMPASS tokamak is planned to be calculated from high-resolution spectroscopic measurements of the carbon III triplet at 465 nm. For these purposes a two-grating spectrometer was provided by Instituto de Plasmas e Fusao Nuclear, Lisboa, Portugal [Gomes et al., 2003; Sokolova et al., 1997].. The light diffracted by the spectrometer is detected by a spectroscopic camera equipped with a CCD chip. In order to interpret the measured C III spectra correctly the system has to be spectrally calibrated. Therefore calibration spectra produced by a Fe hollow cathode lamp are measured. The analysis of the measured calibration spectra provides the calibration function and instrumental line broadening caused by the spectrometer. In this paper the design of the spectrometer is introduced and the calibration procedure is elaborated together with analysis of the calibration spectra, calculation of the calibration function and instrumental width.
Název v anglickém jazyce
Calibration of High-resolution Spectrometer for C III Triplet Measurements
Popis výsledku anglicky
Edge poloidal plasma rotation and ion temperature at the COMPASS tokamak is planned to be calculated from high-resolution spectroscopic measurements of the carbon III triplet at 465 nm. For these purposes a two-grating spectrometer was provided by Instituto de Plasmas e Fusao Nuclear, Lisboa, Portugal [Gomes et al., 2003; Sokolova et al., 1997].. The light diffracted by the spectrometer is detected by a spectroscopic camera equipped with a CCD chip. In order to interpret the measured C III spectra correctly the system has to be spectrally calibrated. Therefore calibration spectra produced by a Fe hollow cathode lamp are measured. The analysis of the measured calibration spectra provides the calibration function and instrumental line broadening caused by the spectrometer. In this paper the design of the spectrometer is introduced and the calibration procedure is elaborated together with analysis of the calibration spectra, calculation of the calibration function and instrumental width.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
BL - Fyzika plasmatu a výboje v plynech
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2015
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
WDS'15 Proceedings of Contributed Papers - Physics
ISBN
978-80-7378-311-2
ISSN
—
e-ISSN
—
Počet stran výsledku
5
Strana od-do
274-278
Název nakladatele
Matfyzpress
Místo vydání
Praha
Místo konání akce
Prague, Czech Republic
Datum konání akce
2. 6. 2015
Typ akce podle státní příslušnosti
EUR - Evropská akce
Kód UT WoS článku
—