Percutaneous Epicardial Pacing Using a Novel Insulated Multi-Electrode Lead
Popis výsledku v původním jazyce
Objectives This study hypothesized that shielded electrodes could capture myocardium without extracardiac stimulation. Background Epicardial cardiac resynchronization therapy (CRT) permits unrestricted electrode positioning. However, this therapy requires surgical placement of device leads and risks unwanted phrenic nerve stimulation. Methods In 6 dog and 5 swine experiments, we used a percutaneous approach to access the epicardial surface of the heart and deployed novel leads housing multiple electrodes with selective insulation. Bipolar pacing thresholds at pre-specified sites were tested to compare electrode threshold data, facing both toward and away from the epicardial surface. Results In 151 paired electrode recordings (70 in 6 dogs; 81 in 5 swine), thresholds facing myocardium were lower than those facing away (median threshold of 0.9 [interquartile range (IQR): 0.4 to 1.6] mA vs. 4.6 [IQR: 2.1 to >10.0] mA, respectively, for dogs, p < 0.0001; and 0.5 [IQR: 0.2 to 1.0] mA
Název v anglickém jazyce
Percutaneous Epicardial Pacing Using a Novel Insulated Multi-Electrode Lead
Popis výsledku anglicky
Objectives This study hypothesized that shielded electrodes could capture myocardium without extracardiac stimulation. Background Epicardial cardiac resynchronization therapy (CRT) permits unrestricted electrode positioning. However, this therapy requires surgical placement of device leads and risks unwanted phrenic nerve stimulation. Methods In 6 dog and 5 swine experiments, we used a percutaneous approach to access the epicardial surface of the heart and deployed novel leads housing multiple electrodes with selective insulation. Bipolar pacing thresholds at pre-specified sites were tested to compare electrode threshold data, facing both toward and away from the epicardial surface. Results In 151 paired electrode recordings (70 in 6 dogs; 81 in 5 swine), thresholds facing myocardium were lower than those facing away (median threshold of 0.9 [interquartile range (IQR): 0.4 to 1.6] mA vs. 4.6 [IQR: 2.1 to >10.0] mA, respectively, for dogs, p < 0.0001; and 0.5 [IQR: 0.2 to 1.0] mA
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
FS - Lékařská zařízení, přístroje a vybavení
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/ED1.100%2F02%2F0123" target="_blank" >ED1.100/02/0123: Fakultní nemocnice u sv. Anny v Brně - Mezinárodní centrum klinického výzkumu (FNUSA - ICRC)</a><br>
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2015
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
JACC: Clinical Electrophysiology
ISSN
2405-500X
e-ISSN
—
Svazek periodika
1
Číslo periodika v rámci svazku
4
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
11
Strana od-do
273-283
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—