Label-free bead-based metallothionein electrochemical immunosensor
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216224%3A14110%2F15%3A00087442" target="_blank" >RIV/00216224:14110/15:00087442 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/00216305:26620/15:PU118478 RIV/62156489:43210/15:43907686
Výsledek na webu
<a href="http://dx.doi.org/10.1002/elps.201500069" target="_blank" >http://dx.doi.org/10.1002/elps.201500069</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1002/elps.201500069" target="_blank" >10.1002/elps.201500069</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Label-free bead-based metallothionein electrochemical immunosensor
Popis výsledku v původním jazyce
A novel microfluidic label-free bead-based metallothionein immunosensors was designed. To the surface of superparamagnetic agarose beads coated with protein A, polyclonal chicken IgY specifically recognizing metallothionein (MT) were immobilized via rabbit IgG. The Brdicka reaction was used for metallothionein detection in a microfluidic printed 3D chip. Assembled chip consisted of a single copper wire coated with a thin layer of amalgam as working electrode. Optimization of MT detection using designed microfluidic chip was performed in stationary system as well as in the flow arrangement at various flow rates (0-1800 mikroL/min). In stationary arrangement it is possible to detect MT concentrations up to 30 ng/mL level, flow arrangement allows reliable detection of even lower concentration (12.5 ng/mL). The assembled miniature flow chip was subsequently tested for the detection of MT elevated levels (at approx. level 100 mikrog/mL) in real samples of patients with cancer.
Název v anglickém jazyce
Label-free bead-based metallothionein electrochemical immunosensor
Popis výsledku anglicky
A novel microfluidic label-free bead-based metallothionein immunosensors was designed. To the surface of superparamagnetic agarose beads coated with protein A, polyclonal chicken IgY specifically recognizing metallothionein (MT) were immobilized via rabbit IgG. The Brdicka reaction was used for metallothionein detection in a microfluidic printed 3D chip. Assembled chip consisted of a single copper wire coated with a thin layer of amalgam as working electrode. Optimization of MT detection using designed microfluidic chip was performed in stationary system as well as in the flow arrangement at various flow rates (0-1800 mikroL/min). In stationary arrangement it is possible to detect MT concentrations up to 30 ng/mL level, flow arrangement allows reliable detection of even lower concentration (12.5 ng/mL). The assembled miniature flow chip was subsequently tested for the detection of MT elevated levels (at approx. level 100 mikrog/mL) in real samples of patients with cancer.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CE - Biochemie
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2015
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Electrophoresis
ISSN
0173-0835
e-ISSN
—
Svazek periodika
36
Číslo periodika v rámci svazku
16
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
11
Strana od-do
1894-1904
Kód UT WoS článku
000360055300014
EID výsledku v databázi Scopus
—