Main symptoms found in patients with same diseases as for example COVID-19 is febrile. The infrared thermography (IRT) represents a fast measurement in case of screening in public places. One of the limitations of IRT is the resolution of sensor, which has close connection with the distance between camera and ROI. To maximize the effectivity of resolution of the camera is to reduce the distance from the object. The aim of presented study showed the possibility how to protect the camera or medical staff that operates the device against potential infection or contamination from the person with infection. Two protective foils of different thickness (40microm; 9microm) were tested as a barrier between the IRT and the ROI (black body model and human face). Even though the results have shown that the transparent foils decrease linearly the measured value of the temperature, it can be used as a protective barrier between IRT and the object if an appropriate recalculation is done during analysis of IRT images. Results are acceptable in the case of 9microm foil especially. The authors see this possibility as a minor concession from IRT standards but as a great help in health protection. The transparent foil can be used as protective barrier of the infrared camera.
Název v anglickém jazyce
Protective aspects in contactless infrared thermography fever screening
Popis výsledku anglicky
Main symptoms found in patients with same diseases as for example COVID-19 is febrile. The infrared thermography (IRT) represents a fast measurement in case of screening in public places. One of the limitations of IRT is the resolution of sensor, which has close connection with the distance between camera and ROI. To maximize the effectivity of resolution of the camera is to reduce the distance from the object. The aim of presented study showed the possibility how to protect the camera or medical staff that operates the device against potential infection or contamination from the person with infection. Two protective foils of different thickness (40microm; 9microm) were tested as a barrier between the IRT and the ROI (black body model and human face). Even though the results have shown that the transparent foils decrease linearly the measured value of the temperature, it can be used as a protective barrier between IRT and the object if an appropriate recalculation is done during analysis of IRT images. Results are acceptable in the case of 9microm foil especially. The authors see this possibility as a minor concession from IRT standards but as a great help in health protection. The transparent foil can be used as protective barrier of the infrared camera.
Klasifikace
Druh
J<sub>SC</sub> - Článek v periodiku v databázi SCOPUS
CEP obor
—
OECD FORD obor
20601 - Medical engineering
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2020
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Lékař a technika
ISSN
0301-5491
e-ISSN
2336-5552
Svazek periodika
50
Číslo periodika v rámci svazku
3
Stát vydavatele periodika
CZ - Česká republika
Počet stran výsledku
6
Strana od-do
95-100