Development of piezoelectric immunosensors for measurement of albuminuria
Popis výsledku v původním jazyce
The development of piezoelectric immunosensors for human serum albumin (HSA) is reported. The piezoelectric crystals were modified either with monoclonal antibody AL-01 (direct assay) or with HSA (competitive assay). Measurements were carried out in theflow-through mode. Affinity interaction between albumin and the antibody was characterised. With immobilised antibody and HSA in solution, the kinetic association rate constant ka was 18100 l mol-1 s-1 and the dissociation constant kd was 0.00369 s-1. For the opposite arrangement (immobilised HSA), a slower dissociation was observed, kd was 0.00085 s-1. A competitive assay for HSA was developed with working range of 1 to 5 ug/ml and a total time for one analysis equal to 17 min. Samples of urine were analysed after tenfold dilution. The developed system was successfully evaluated on real samples from diabetic patients and the obtained results correlated well with the standard reflectometric assay of proteins in urine.
Název v anglickém jazyce
Development of piezoelectric immunosensors for measurement of albuminuria
Popis výsledku anglicky
The development of piezoelectric immunosensors for human serum albumin (HSA) is reported. The piezoelectric crystals were modified either with monoclonal antibody AL-01 (direct assay) or with HSA (competitive assay). Measurements were carried out in theflow-through mode. Affinity interaction between albumin and the antibody was characterised. With immobilised antibody and HSA in solution, the kinetic association rate constant ka was 18100 l mol-1 s-1 and the dissociation constant kd was 0.00369 s-1. For the opposite arrangement (immobilised HSA), a slower dissociation was observed, kd was 0.00085 s-1. A competitive assay for HSA was developed with working range of 1 to 5 ug/ml and a total time for one analysis equal to 17 min. Samples of urine were analysed after tenfold dilution. The developed system was successfully evaluated on real samples from diabetic patients and the obtained results correlated well with the standard reflectometric assay of proteins in urine.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CE - Biochemie
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2001
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Talanta
ISSN
0039-9140
e-ISSN
—
Svazek periodika
55
Číslo periodika v rámci svazku
11
Stát vydavatele periodika
NL - Nizozemsko
Počet stran výsledku
9
Strana od-do
831
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—