Immunosensor for the measurement of human serum albumin in urine based on the Spreeta surface plasmon resonance sensor
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216224%3A14310%2F03%3A00008617" target="_blank" >RIV/00216224:14310/03:00008617 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Immunosensor for the measurement of human serum albumin in urine based on the Spreeta surface plasmon resonance sensor
Popis výsledku v původním jazyce
The application of SPR for the measurement of concentration of human serum albumin (HSA) in urine was studied using the compact integrated SPR sensing system Spreeta. HSA was immobilized via cystamine and glutaraldehyde onto the gold sensing area and a competitive assay for HSA was developed using limited amount of the monoclonal antibody AL-01 in solution. Measurements were carried out in the flow-through mode and the interaction between immobilized HSA and antibody was observed in real time. To obtainreproducible results, different conditions of the measurement (method of immobilization of HSA, data evaluation, concentration of antibody, regeneration procedure) were tested. The calibration curve for clinically relevant concentrations of HSA in urinesamples was constructed using 300-times diluted antibody in the form of ascites fluid. The measuring range was between 0.1 and 5 mg/l of HSA, the sensing surface was successfully regenerated and suitable for more than 20 assays. The deve
Název v anglickém jazyce
Immunosensor for the measurement of human serum albumin in urine based on the Spreeta surface plasmon resonance sensor
Popis výsledku anglicky
The application of SPR for the measurement of concentration of human serum albumin (HSA) in urine was studied using the compact integrated SPR sensing system Spreeta. HSA was immobilized via cystamine and glutaraldehyde onto the gold sensing area and a competitive assay for HSA was developed using limited amount of the monoclonal antibody AL-01 in solution. Measurements were carried out in the flow-through mode and the interaction between immobilized HSA and antibody was observed in real time. To obtainreproducible results, different conditions of the measurement (method of immobilization of HSA, data evaluation, concentration of antibody, regeneration procedure) were tested. The calibration curve for clinically relevant concentrations of HSA in urinesamples was constructed using 300-times diluted antibody in the form of ascites fluid. The measuring range was between 0.1 and 5 mg/l of HSA, the sensing surface was successfully regenerated and suitable for more than 20 assays. The deve
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CE - Biochemie
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/OC%20518.30" target="_blank" >OC 518.30: Sledování interakcí biomakromolekul s biovrstvami v různých konformačních stavech pomocí piezoelektrických biosensorů</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2003
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Supramolecular Chemistry
ISSN
1061-0278
e-ISSN
—
Svazek periodika
15
Číslo periodika v rámci svazku
2
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
7
Strana od-do
109-115
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—