Comparison of five integrative samplers in laboratory for the monitoring of indicator and dioxin-like polychlorinated biphenyls in water
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216224%3A14310%2F14%3A00079311" target="_blank" >RIV/00216224:14310/14:00079311 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1016/j.chemosphere.2013.09.011" target="_blank" >http://dx.doi.org/10.1016/j.chemosphere.2013.09.011</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.chemosphere.2013.09.011" target="_blank" >10.1016/j.chemosphere.2013.09.011</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Comparison of five integrative samplers in laboratory for the monitoring of indicator and dioxin-like polychlorinated biphenyls in water
Popis výsledku v původním jazyce
This study aimed at evaluating and comparing five integrative samplers for the monitoring of indicator and dioxin-like polychlorinated biphenyls (PCBs) in water: semi-permeable membrane device (SPMD), silicone rubber, low-density polyethylene (LOPE) strip, Chemcatcher and a continuous-flow integrative sampler (CFIS). These samplers were spiked with performance reference compounds (PRCs) and then simultaneously exposed under constant agitation and temperature in a 200 L stainless steel tank for periods ranging from one day to three months. A constant PCB concentration of about 1 ng.L-1 was achieved by immersing a large amount of silicone rubber sheets ("dosing sheets") spiked with the target PCBs. The uptake of PCBs in the five samplers showed overall good repeatability and their accumulation was linear with time. The samplers SPMD, silicone rubber and LOPE strip were the most promising in terms of achieving low limits of quantification.
Název v anglickém jazyce
Comparison of five integrative samplers in laboratory for the monitoring of indicator and dioxin-like polychlorinated biphenyls in water
Popis výsledku anglicky
This study aimed at evaluating and comparing five integrative samplers for the monitoring of indicator and dioxin-like polychlorinated biphenyls (PCBs) in water: semi-permeable membrane device (SPMD), silicone rubber, low-density polyethylene (LOPE) strip, Chemcatcher and a continuous-flow integrative sampler (CFIS). These samplers were spiked with performance reference compounds (PRCs) and then simultaneously exposed under constant agitation and temperature in a 200 L stainless steel tank for periods ranging from one day to three months. A constant PCB concentration of about 1 ng.L-1 was achieved by immersing a large amount of silicone rubber sheets ("dosing sheets") spiked with the target PCBs. The uptake of PCBs in the five samplers showed overall good repeatability and their accumulation was linear with time. The samplers SPMD, silicone rubber and LOPE strip were the most promising in terms of achieving low limits of quantification.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
DJ - Znečištění a kontrola vody
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2014
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Chemosphere
ISSN
0045-6535
e-ISSN
—
Svazek periodika
98
Číslo periodika v rámci svazku
MAR
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
10
Strana od-do
18-27
Kód UT WoS článku
000331679600002
EID výsledku v databázi Scopus
—