Automated micro axial tomography of cell nuclei after specific labelling by fluorescence in situ hybridisation
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216224%3A14330%2F02%3A00006215" target="_blank" >RIV/00216224:14330/02:00006215 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/68081707:_____/02:17033133
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Automated micro axial tomography of cell nuclei after specific labelling by fluorescence in situ hybridisation
Popis výsledku v původním jazyce
Micro axial tomography provides a good means for microscopic image acquisition of cells or sub-cellular components like cell nuclei with an improved resolution, because shortcomings of spatial resolution anisotropy in optical microscopy can be overcome.Thus, spatial information of the object can be obtained without the necessity of confocal imaging. Since the very early developments of micro axial tomography a considerable drawback of this method was a complicated image acquisition and processing procedure that requires much operator time. In order to solve this problem the Heidelberg 2pi-tilting device has been mounted to the Brno high-resolution cytometer as an attempt to bring together advanced microscopy and fast automated computer image acquisition and analysis. A special software module that drives all hardware components required for automated micro axial tomography and performs image acquisition and processing has been developed. Also a method for the preparation of cell nucle
Název v anglickém jazyce
Automated micro axial tomography of cell nuclei after specific labelling by fluorescence in situ hybridisation
Popis výsledku anglicky
Micro axial tomography provides a good means for microscopic image acquisition of cells or sub-cellular components like cell nuclei with an improved resolution, because shortcomings of spatial resolution anisotropy in optical microscopy can be overcome.Thus, spatial information of the object can be obtained without the necessity of confocal imaging. Since the very early developments of micro axial tomography a considerable drawback of this method was a complicated image acquisition and processing procedure that requires much operator time. In order to solve this problem the Heidelberg 2pi-tilting device has been mounted to the Brno high-resolution cytometer as an attempt to bring together advanced microscopy and fast automated computer image acquisition and analysis. A special software module that drives all hardware components required for automated micro axial tomography and performs image acquisition and processing has been developed. Also a method for the preparation of cell nucle
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
JD - Využití počítačů, robotika a její aplikace
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2002
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Micron
ISSN
0968-4328
e-ISSN
—
Svazek periodika
33
Číslo periodika v rámci svazku
7
Stát vydavatele periodika
NL - Nizozemsko
Počet stran výsledku
11
Strana od-do
655
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—