Optimization of two-dimensional gradient liquid chromatography separations
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216275%3A25310%2F09%3A00008420" target="_blank" >RIV/00216275:25310/09:00008420 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Optimization of two-dimensional gradient liquid chromatography separations
Popis výsledku v původním jazyce
An almost orthogonal comprehensive two-dimensional liquid chromatogramy was developed for the separation of phenolic and flavone natural antioxidants by using combinations of a polyethylene glycol silica micro-column in the first dimension and a porous-shell fused-core C18 column in the second dimension, both in the reversed-phase mode. System ortogonality was improved using parallel gradients of acetonitrile in buffered mobile phase. A new approach was proposed to optimize matching segmented gradient profiles in the two dimensions. An algorithm was developed for automatic corrections of the shifts in retention in the second dimension induced by the parallel two-dimensional gradient operation technique. Using the porous-shell C18 column in the second dimension at elevated temperature (60 °C) and high pressure (480 bar) with optimized segmented profiles of the parallel gradients in the two dimensions, the overall separation time for comprehensive LCxLC was reduced to 30 min.
Název v anglickém jazyce
Optimization of two-dimensional gradient liquid chromatography separations
Popis výsledku anglicky
An almost orthogonal comprehensive two-dimensional liquid chromatogramy was developed for the separation of phenolic and flavone natural antioxidants by using combinations of a polyethylene glycol silica micro-column in the first dimension and a porous-shell fused-core C18 column in the second dimension, both in the reversed-phase mode. System ortogonality was improved using parallel gradients of acetonitrile in buffered mobile phase. A new approach was proposed to optimize matching segmented gradient profiles in the two dimensions. An algorithm was developed for automatic corrections of the shifts in retention in the second dimension induced by the parallel two-dimensional gradient operation technique. Using the porous-shell C18 column in the second dimension at elevated temperature (60 °C) and high pressure (480 bar) with optimized segmented profiles of the parallel gradients in the two dimensions, the overall separation time for comprehensive LCxLC was reduced to 30 min.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CB - Analytická chemie, separace
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GA203%2F07%2F0641" target="_blank" >GA203/07/0641: Nové analytické postupy v kombinovaných fázových systémech s vysokou separační kapacitou</a><br>
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2009
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Journal of Chromatography A
ISSN
0021-9673
e-ISSN
—
Svazek periodika
1216
Číslo periodika v rámci svazku
16
Stát vydavatele periodika
NL - Nizozemsko
Počet stran výsledku
15
Strana od-do
—
Kód UT WoS článku
—