Voltammetric determination of leucovorin using silver solid amalgam electrode
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216275%3A25310%2F12%3A39895237" target="_blank" >RIV/00216275:25310/12:39895237 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/61388955:_____/12:00375916
Výsledek na webu
<a href="http://dx.doi.org/10.1016/j.electacta.2011.11.071" target="_blank" >http://dx.doi.org/10.1016/j.electacta.2011.11.071</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.electacta.2011.11.071" target="_blank" >10.1016/j.electacta.2011.11.071</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Voltammetric determination of leucovorin using silver solid amalgam electrode
Popis výsledku v původním jazyce
The voltammetric behavior of leucovorin (LV), as an important and stable reduced metabolite of folic acid, has been investigated on polished (p-AgSAE) and mercury meniscus modified (m-AgSAE) silver solid amalgam electrodes in this paper. LV, after its reverse electrochemical oxidation, can be reduced in two reduction steps in an acidic medium on both electrodes used. The first reduction peak (E-p = -400 mV) can be used for determination of LV. Differential Pulse Voltammetry (DPV) was applied for monitoring voltammetric behavior of LV on both working electrodes. Elimination voltammetry with linear scan was applied for elucidation of the reaction mechanisms. The limits of detection were calculated as 2.2 x 10(-8) mol L-1 (t(acc) = 60 s) for m-AgSAE and 5.0 x 10(-8) mol L-1 (t(acc) = 100 s) for p-AgSAE. The proposed method was applied for LV determination in the pharmaceutical preparation which has been commonly used during methotrexate therapy. (C) 2011 Published by Elsevier Ltd.
Název v anglickém jazyce
Voltammetric determination of leucovorin using silver solid amalgam electrode
Popis výsledku anglicky
The voltammetric behavior of leucovorin (LV), as an important and stable reduced metabolite of folic acid, has been investigated on polished (p-AgSAE) and mercury meniscus modified (m-AgSAE) silver solid amalgam electrodes in this paper. LV, after its reverse electrochemical oxidation, can be reduced in two reduction steps in an acidic medium on both electrodes used. The first reduction peak (E-p = -400 mV) can be used for determination of LV. Differential Pulse Voltammetry (DPV) was applied for monitoring voltammetric behavior of LV on both working electrodes. Elimination voltammetry with linear scan was applied for elucidation of the reaction mechanisms. The limits of detection were calculated as 2.2 x 10(-8) mol L-1 (t(acc) = 60 s) for m-AgSAE and 5.0 x 10(-8) mol L-1 (t(acc) = 100 s) for p-AgSAE. The proposed method was applied for LV determination in the pharmaceutical preparation which has been commonly used during methotrexate therapy. (C) 2011 Published by Elsevier Ltd.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CG - Elektrochemie
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2012
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Electrochimica Acta
ISSN
0013-4686
e-ISSN
—
Svazek periodika
60
Číslo periodika v rámci svazku
January
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
9
Strana od-do
375-383
Kód UT WoS článku
000300142100052