In this paper, pretreated pencil graphite electrode (Pre-PGE) was used for the determination of insulin by anodic stripping differential pulse voltammetry (ASDPV). The influences of the pretreatment procedures, pH values, accumulation potential and timeon the signal enhancement of insulin were optimized, and a novel electrochemical method for the determination of insulin was described. The currents obtained from ASDPV measurements at optimum conditions (phosphate buffer solution pH 7.0), accumulation time (60 seconds), applied potential 400 mV vs. Ag/AgCl/3.0 mol L-1 KCl) were linearly correlated with the concentration of insulin. Calibration curve was obtained for insulin concentrations in the range of 5.0-200.0x10-9 mol L-1 .The limit of detection was found to be 2.0x10-9 mol L-1.
Název v anglickém jazyce
Determination of Insulin Using a Pretreated Pencil Graphite Electrode
Popis výsledku anglicky
In this paper, pretreated pencil graphite electrode (Pre-PGE) was used for the determination of insulin by anodic stripping differential pulse voltammetry (ASDPV). The influences of the pretreatment procedures, pH values, accumulation potential and timeon the signal enhancement of insulin were optimized, and a novel electrochemical method for the determination of insulin was described. The currents obtained from ASDPV measurements at optimum conditions (phosphate buffer solution pH 7.0), accumulation time (60 seconds), applied potential 400 mV vs. Ag/AgCl/3.0 mol L-1 KCl) were linearly correlated with the concentration of insulin. Calibration curve was obtained for insulin concentrations in the range of 5.0-200.0x10-9 mol L-1 .The limit of detection was found to be 2.0x10-9 mol L-1.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
CG - Elektrochemie
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach<br>I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2014
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Sensing in Electroanalysis Vol. 8
ISBN
978-80-7395-782-7
ISSN
—
e-ISSN
—
Počet stran výsledku
9
Strana od-do
305-313
Název nakladatele
Univerzita Pardubice
Místo vydání
Pardubice
Místo konání akce
Pardubice
Datum konání akce
13. 11. 2013
Typ akce podle státní příslušnosti
EUR - Evropská akce
Kód UT WoS článku
—