Simple and Rapid Determination of Ethanol Content in Beer Using an Amperometric Biosensor
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216275%3A25310%2F15%3A39900647" target="_blank" >RIV/00216275:25310/15:39900647 - isvavai.cz</a>
Výsledek na webu
<a href="http://www.mdpi.com/2227-9040/3/2/169" target="_blank" >http://www.mdpi.com/2227-9040/3/2/169</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.3390/chemosensors3020169" target="_blank" >10.3390/chemosensors3020169</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Simple and Rapid Determination of Ethanol Content in Beer Using an Amperometric Biosensor
Popis výsledku v původním jazyce
An alcohol dehydrogenase-based biosensor was prepared and tested for its use to determine ethanol in beer. The biosensor is based on a screen-printed carbon electrode (SPCE) modified by rhodium dioxide and immobilized with a biocatalytic layer containingthe enzyme. Function of the enzyme biosensor was tested in model ethanol samples, in which it showed a linear range of 15-120 g.L-1 with a detection limit of 3.3 g.L-1 (established as 3 sigma) and response time of 19 s. In a potential window from -0.2 to +0.45 V, interferences of both ascorbic and uric acids were negligible. Several types of marketed beers of Czech provenance were selected and subjected to measurements under optimized conditions but without any pretreatment of real samples. When compared with the reference method (gas chromatography), the results were in quite good agreement for beers of the pale lager type but higher contents of ethanol were indicated in the samples of dark lager beers.
Název v anglickém jazyce
Simple and Rapid Determination of Ethanol Content in Beer Using an Amperometric Biosensor
Popis výsledku anglicky
An alcohol dehydrogenase-based biosensor was prepared and tested for its use to determine ethanol in beer. The biosensor is based on a screen-printed carbon electrode (SPCE) modified by rhodium dioxide and immobilized with a biocatalytic layer containingthe enzyme. Function of the enzyme biosensor was tested in model ethanol samples, in which it showed a linear range of 15-120 g.L-1 with a detection limit of 3.3 g.L-1 (established as 3 sigma) and response time of 19 s. In a potential window from -0.2 to +0.45 V, interferences of both ascorbic and uric acids were negligible. Several types of marketed beers of Czech provenance were selected and subjected to measurements under optimized conditions but without any pretreatment of real samples. When compared with the reference method (gas chromatography), the results were in quite good agreement for beers of the pale lager type but higher contents of ethanol were indicated in the samples of dark lager beers.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CG - Elektrochemie
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2015
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Chemosensors
ISSN
2227-9040
e-ISSN
—
Svazek periodika
3
Číslo periodika v rámci svazku
2
Stát vydavatele periodika
CH - Švýcarská konfederace
Počet stran výsledku
9
Strana od-do
169-177
Kód UT WoS článku
000362073300004
EID výsledku v databázi Scopus
—