Carbon microparticles as a physical carrier for ETV-ICP-MS
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216275%3A25310%2F19%3A39914891" target="_blank" >RIV/00216275:25310/19:39914891 - isvavai.cz</a>
Výsledek na webu
<a href="https://www.sciencedirect.com/science/article/pii/S0039914019306708" target="_blank" >https://www.sciencedirect.com/science/article/pii/S0039914019306708</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.talanta.2019.06.045" target="_blank" >10.1016/j.talanta.2019.06.045</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Carbon microparticles as a physical carrier for ETV-ICP-MS
Popis výsledku v původním jazyce
In this study, carbon microparticles (CMs) as a physical carrier to enhance the analyte transport efficiency for inductively coupled plasma mass spectrometry connected with electrothermal vaporization (ETV-ICP-MS) are proposed. Carbon microparticles mixed with samples or calibration standard solutions were dosed as a slurry into the graphite furnace. The optimization of working conditions was done for the standard solution 0.5 mu g L-1 of Au and Tl. The pyrolysis and vaporization temperatures were 500 and 2700 degrees C for Au and 400 and 1900 degrees C for Tl. The optimized CMs concentration was 1 g L-1 for Au and 2.5 g L-1 for Tl. For the quantification, external calibration standard solutions were used. The result obtained for Au in the digested CRM GBW 07601 (the informatory value 2.1 +/- 2 mu g kg(-1)) was 2.2 +/- 0.1 mu g kg(-1). The result for Tl in the digested CRM BCR 679 (the certified value 3 +/- 0.3 mu g kg(-1)) was 3.2 +/- 0.2 mu g kg(-1) and in the slurry 2.7 +/- 0.1 mu g kg(-1). The result for Tl in the CRM GBW 10052 (the certified value 57 +/- 11 mu g kg(-1)) was 51 +/- 3 mu g kg(-1). The instrumental limits of detection were 0.016 ng L-1 for Au and 0.026 ng L-1 for Tl. The recoveries and repeatabilities measured on calibration standards were in the range 99-100% and 0.2-2.3% for Au and 100-111% and 2.9-6.7% for Tl.
Název v anglickém jazyce
Carbon microparticles as a physical carrier for ETV-ICP-MS
Popis výsledku anglicky
In this study, carbon microparticles (CMs) as a physical carrier to enhance the analyte transport efficiency for inductively coupled plasma mass spectrometry connected with electrothermal vaporization (ETV-ICP-MS) are proposed. Carbon microparticles mixed with samples or calibration standard solutions were dosed as a slurry into the graphite furnace. The optimization of working conditions was done for the standard solution 0.5 mu g L-1 of Au and Tl. The pyrolysis and vaporization temperatures were 500 and 2700 degrees C for Au and 400 and 1900 degrees C for Tl. The optimized CMs concentration was 1 g L-1 for Au and 2.5 g L-1 for Tl. For the quantification, external calibration standard solutions were used. The result obtained for Au in the digested CRM GBW 07601 (the informatory value 2.1 +/- 2 mu g kg(-1)) was 2.2 +/- 0.1 mu g kg(-1). The result for Tl in the digested CRM BCR 679 (the certified value 3 +/- 0.3 mu g kg(-1)) was 3.2 +/- 0.2 mu g kg(-1) and in the slurry 2.7 +/- 0.1 mu g kg(-1). The result for Tl in the CRM GBW 10052 (the certified value 57 +/- 11 mu g kg(-1)) was 51 +/- 3 mu g kg(-1). The instrumental limits of detection were 0.016 ng L-1 for Au and 0.026 ng L-1 for Tl. The recoveries and repeatabilities measured on calibration standards were in the range 99-100% and 0.2-2.3% for Au and 100-111% and 2.9-6.7% for Tl.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
10406 - Analytical chemistry
Návaznosti výsledku
Projekt
<a href="/cs/project/EF16_013%2F0001829" target="_blank" >EF16_013/0001829: Modernizace a upgrade infrastruktury CEMNAT</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2019
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Talanta
ISSN
0039-9140
e-ISSN
—
Svazek periodika
204
Číslo periodika v rámci svazku
November
Stát vydavatele periodika
NL - Nizozemsko
Počet stran výsledku
6
Strana od-do
555-560
Kód UT WoS článku
000480664200068
EID výsledku v databázi Scopus
—