Comparison of the computational and experimental modelling results of the hip joint endoprosthesis behaviour under pressing on the stem
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26210%2F03%3APU37903" target="_blank" >RIV/00216305:26210/03:PU37903 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/61388998:_____/03:51030130
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Comparison of the computational and experimental modelling results of the hip joint endoprosthesis behaviour under pressing on the stem
Popis výsledku v původním jazyce
This paper is concerned with the comparison of the computational and experimental results of the hip joint endoprosthesis ceramic head and stem cone self-adhesive connection. During the experiment according the ISO 7206-5 there were measured the tangential strain on the lower part of the head's sphere and vertically head displacement. The system FEM ANSYS was used for the computational modelling of the system deformation and stresses. The measured shape deflections from ideal cone shapes (from IMS UMPIRRE device) were implemented to the FEM model. The comparison of the computational and experimental data were realised for variants with the smooth and teethless stem.
Název v anglickém jazyce
Comparison of the computational and experimental modelling results of the hip joint endoprosthesis behaviour under pressing on the stem
Popis výsledku anglicky
This paper is concerned with the comparison of the computational and experimental results of the hip joint endoprosthesis ceramic head and stem cone self-adhesive connection. During the experiment according the ISO 7206-5 there were measured the tangential strain on the lower part of the head's sphere and vertically head displacement. The system FEM ANSYS was used for the computational modelling of the system deformation and stresses. The measured shape deflections from ideal cone shapes (from IMS UMPIRRE device) were implemented to the FEM model. The comparison of the computational and experimental data were realised for variants with the smooth and teethless stem.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
FS - Lékařská zařízení, přístroje a vybavení
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GP101%2F01%2FP039" target="_blank" >GP101/01/P039: Stochastický přístup k posouzení spolehlivosti keramických hlavic totální kyčelní endoprotézy</a><br>
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2003
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Experimental Stress Analysis 2003
ISBN
80-214-2314-5
ISSN
—
e-ISSN
—
Počet stran výsledku
2
Strana od-do
27-28
Název nakladatele
Brno University of Technology, FME, Institute of Mechanics of Solids
Místo vydání
Milovy
Místo konání akce
Milovy
Datum konání akce
3. 6. 2003
Typ akce podle státní příslušnosti
EUR - Evropská akce