Evaluating Different Shapes of Cranial Fixation Mini-plates Using Finite Element Method
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26210%2F17%3APU123859" target="_blank" >RIV/00216305:26210/17:PU123859 - isvavai.cz</a>
Výsledek na webu
<a href="https://link.springer.com/chapter/10.1007/978-981-10-5122-7_187" target="_blank" >https://link.springer.com/chapter/10.1007/978-981-10-5122-7_187</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1007/978-981-10-5122-7_187" target="_blank" >10.1007/978-981-10-5122-7_187</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Evaluating Different Shapes of Cranial Fixation Mini-plates Using Finite Element Method
Popis výsledku v původním jazyce
Medical grade 3D printing offers the possibility to manufacture patient-specific implants to treat cranial defects. The performance of the implant assembly depends on many factors, such as material, thickness, size and manufacturing accuracy. A significant factor in the stability and success of the assembly is the fixation method. Cranial implants are usually fixed to the skull by means of mini-plates. Biomechanical assessment of fixing the implant to the skull might be helpful not only for mini-plate design but might be beneficial also for the surgeons. In this study, four different mini-plate designs were analyzed and compared based on the stress-strain analysis of one cranial implant fixed at three locations by mini-plates. Computational simulations were done using Finite Element Method.
Název v anglickém jazyce
Evaluating Different Shapes of Cranial Fixation Mini-plates Using Finite Element Method
Popis výsledku anglicky
Medical grade 3D printing offers the possibility to manufacture patient-specific implants to treat cranial defects. The performance of the implant assembly depends on many factors, such as material, thickness, size and manufacturing accuracy. A significant factor in the stability and success of the assembly is the fixation method. Cranial implants are usually fixed to the skull by means of mini-plates. Biomechanical assessment of fixing the implant to the skull might be helpful not only for mini-plate design but might be beneficial also for the surgeons. In this study, four different mini-plate designs were analyzed and compared based on the stress-strain analysis of one cranial implant fixed at three locations by mini-plates. Computational simulations were done using Finite Element Method.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
—
OECD FORD obor
10610 - Biophysics
Návaznosti výsledku
Projekt
<a href="/cs/project/GA16-08944S" target="_blank" >GA16-08944S: Biomechanická studie interakce kraniomaxilofaciálního implantátu s kostí s využitím personalizovaného přístupu a víceúrovňového výpočtového modelování</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2017
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
EMBEC’17 & NBC’17
ISBN
978-981-10-5122-7
ISSN
1680-0737
e-ISSN
—
Počet stran výsledku
4
Strana od-do
747-750
Název nakladatele
Springer
Místo vydání
Neuveden
Místo konání akce
Tampere
Datum konání akce
11. 6. 2017
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
000449778900187