Wide Range Precise Conductivity Measurement without Cell Replacement
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26220%2F02%3APU29223" target="_blank" >RIV/00216305:26220/02:PU29223 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/10557300:_____/02:#0000006
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Wide Range Precise Conductivity Measurement without Cell Replacement
Popis výsledku v původním jazyce
The paper present new approach in the field of precise electrolytic conductivity measurements with planar sensors. This novel measuring method was suggested for measurement with IDEs thick-film electrodes. Correction characteristics over a wide range ofa specific conductivities were determined from an interface impedance characterisation of the thick-film IDEs. A local minimum of the imaginary part of the interface impedance is used for corrections to get linear responses. The ridge frequency was deterrmined at wide range of measured conductivity. An iteration mode of measurements was suggested to exactly measure the conductivity in the ridge frequency in order to achieve a highly accurate response. The method takes precise conductivity measurements in concentration ranges from 10-6 to 1 M without cell replacement.
Název v anglickém jazyce
Wide Range Precise Conductivity Measurement without Cell Replacement
Popis výsledku anglicky
The paper present new approach in the field of precise electrolytic conductivity measurements with planar sensors. This novel measuring method was suggested for measurement with IDEs thick-film electrodes. Correction characteristics over a wide range ofa specific conductivities were determined from an interface impedance characterisation of the thick-film IDEs. A local minimum of the imaginary part of the interface impedance is used for corrections to get linear responses. The ridge frequency was deterrmined at wide range of measured conductivity. An iteration mode of measurements was suggested to exactly measure the conductivity in the ridge frequency in order to achieve a highly accurate response. The method takes precise conductivity measurements in concentration ranges from 10-6 to 1 M without cell replacement.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JB - Senzory, čidla, měření a regulace
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2002
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Eurosensors XVI, The 16th European Conference on Solid-State Transducers
ISBN
80-01-02576-4
ISSN
—
e-ISSN
—
Počet stran výsledku
2
Strana od-do
602-603
Název nakladatele
Czech Technical UniversityFaculty of Electrical EngineeringDepartment of MeasurementTechnická 2, Prague 6-Dejvice166 27 Czech Republic
Místo vydání
FEL CVUT Praha
Místo konání akce
Praha
Datum konání akce
15. 9. 2002
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—