Zhodnocení MRI artefaktů způsobených kovovými dentálními implantáty a klasifikace použití těchto dentálních materiálů.
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26220%2F06%3APU54971" target="_blank" >RIV/00216305:26220/06:PU54971 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/68081731:_____/06:00048827 RIV/00216208:11110/06:00005393 RIV/00064165:_____/06:00000820
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Evaluation of MRI artifacts caused by metallic dental implants and classification of the dental materials in use
Popis výsledku v původním jazyce
Metallic objects (dental fillings, crowns, fixed bridges etc.), now found in many patients, may become a reason for the patient's ineligibility for MR examination because of safety or image artifact concerns. The severity of the MRI artifacts caused bythese objects depends primarily on the material, shape, position and orientation of the object, MRI method applied, and the location of the tissues of interest. The purpose of this work is to demonstrate the artifacts and describe a method of measuri ing the relevant material properties and the results. A method for the measurement of magnetic susceptibility based on field distortion simulation and field mapping based on MRI is described. Objects made of several materials were tested experimentally and these results were checked against simulated results. It has been found that among the more than 100 dental materials currently in use there are highly paramagnetic materials (Ni-Cr or Co-Cr based alloys), in which B0 inhomogeneity effects
Název v anglickém jazyce
Evaluation of MRI artifacts caused by metallic dental implants and classification of the dental materials in use
Popis výsledku anglicky
Metallic objects (dental fillings, crowns, fixed bridges etc.), now found in many patients, may become a reason for the patient's ineligibility for MR examination because of safety or image artifact concerns. The severity of the MRI artifacts caused bythese objects depends primarily on the material, shape, position and orientation of the object, MRI method applied, and the location of the tissues of interest. The purpose of this work is to demonstrate the artifacts and describe a method of measuri ing the relevant material properties and the results. A method for the measurement of magnetic susceptibility based on field distortion simulation and field mapping based on MRI is described. Objects made of several materials were tested experimentally and these results were checked against simulated results. It has been found that among the more than 100 dental materials currently in use there are highly paramagnetic materials (Ni-Cr or Co-Cr based alloys), in which B0 inhomogeneity effects
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
JA - Elektronika a optoelektronika, elektrotechnika
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/NR8110" target="_blank" >NR8110: Dentální slitiny a jejich kompatibilita s vyšetřením metodou magnetické rezonance</a><br>
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2006
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Measurement Science Review
ISSN
1335-8871
e-ISSN
—
Svazek periodika
6
Číslo periodika v rámci svazku
2
Stát vydavatele periodika
SK - Slovenská republika
Počet stran výsledku
4
Strana od-do
24-27
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—