The determination of function G and air ion mobility spectrum in an aspiration condenser with segmented innerelectrode
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26220%2F12%3APU98489" target="_blank" >RIV/00216305:26220/12:PU98489 - isvavai.cz</a>
Výsledek na webu
<a href="http://piers.org/piersproceedings/piers2012KualalumpurProc.php?searchname=Mikulka" target="_blank" >http://piers.org/piersproceedings/piers2012KualalumpurProc.php?searchname=Mikulka</a>
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
The determination of function G and air ion mobility spectrum in an aspiration condenser with segmented innerelectrode
Popis výsledku v původním jazyce
It was confirmed that light air ions have positive influence on human health. For its appraisal it is necessary to know the concentration of air ions and air ion mobility spectrum. When we measure spectrum of air ions by the standard type of the gerdientube with non-segmented electrode we must change voltage in the time. The disadvantage of this method is impossibility of real-time measurement air ion mobility spectrum because it is time-consuming. If we use aspiration condenser with segmented inner electrode it is possible determination air ion mobility spectrum from current individual segments inner electrode. This is problem because numeric modeling of the electrostatic field inside the aspiration condenser using FEM is necessary and in the next step the determination trajectory of air ion in aspiration condenser. From this result it is possible to determine function G. This function is used for determination of air ion mobility spectrum. This algorithm is showed in this paper.
Název v anglickém jazyce
The determination of function G and air ion mobility spectrum in an aspiration condenser with segmented innerelectrode
Popis výsledku anglicky
It was confirmed that light air ions have positive influence on human health. For its appraisal it is necessary to know the concentration of air ions and air ion mobility spectrum. When we measure spectrum of air ions by the standard type of the gerdientube with non-segmented electrode we must change voltage in the time. The disadvantage of this method is impossibility of real-time measurement air ion mobility spectrum because it is time-consuming. If we use aspiration condenser with segmented inner electrode it is possible determination air ion mobility spectrum from current individual segments inner electrode. This is problem because numeric modeling of the electrostatic field inside the aspiration condenser using FEM is necessary and in the next step the determination trajectory of air ion in aspiration condenser. From this result it is possible to determine function G. This function is used for determination of air ion mobility spectrum. This algorithm is showed in this paper.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JA - Elektronika a optoelektronika, elektrotechnika
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GA102%2F09%2F0314" target="_blank" >GA102/09/0314: Studium vlastností metamateriálů a mikrovlnných struktur s využitím šumové spektroskopie a magnetické rezonance</a><br>
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2012
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Proceedings of PIERS 2012 in Kuala Lumpur
ISBN
978-1-934142-20-2
ISSN
—
e-ISSN
—
Počet stran výsledku
5
Strana od-do
374-379
Název nakladatele
The Electromagnetic Academy
Místo vydání
Cambridge
Místo konání akce
Kuala Lumpur
Datum konání akce
27. 3. 2012
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—