Calibration of the apparatus for 3D magnetic measurement for EIT
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26220%2F12%3APU98490" target="_blank" >RIV/00216305:26220/12:PU98490 - isvavai.cz</a>
Výsledek na webu
<a href="http://piers.org/piersproceedings/piers2012KualalumpurProc.php?searchname=Mikulka" target="_blank" >http://piers.org/piersproceedings/piers2012KualalumpurProc.php?searchname=Mikulka</a>
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Calibration of the apparatus for 3D magnetic measurement for EIT
Popis výsledku v původním jazyce
The paper deals about result measurement for designed apparatus 3D Teslameter magnetic field. This apparatus allows measure direct or alternating magnetic field in environment with strong electromagnetic disturbance. In the case of measure alternating magnetic field is possible measure signal with lower amplitude than electromagnetic disturbance, concrete S/N <0.01. The measurement was performing in workplace for mapping magnetic field. In the designed 3D magnetic field meter, Hall probes are used for the measurement of magnetic flux density; also, the influence of Earths magnetic field is compensated. The measurement result was compared with simulation results in program MATLAB. In this paper are discussing differences between measurement and simulation results and determine source systematic error. The magnetic field map is measured using the calibration 3D meter and is used for the reconstruction of material properties of the measured sample.
Název v anglickém jazyce
Calibration of the apparatus for 3D magnetic measurement for EIT
Popis výsledku anglicky
The paper deals about result measurement for designed apparatus 3D Teslameter magnetic field. This apparatus allows measure direct or alternating magnetic field in environment with strong electromagnetic disturbance. In the case of measure alternating magnetic field is possible measure signal with lower amplitude than electromagnetic disturbance, concrete S/N <0.01. The measurement was performing in workplace for mapping magnetic field. In the designed 3D magnetic field meter, Hall probes are used for the measurement of magnetic flux density; also, the influence of Earths magnetic field is compensated. The measurement result was compared with simulation results in program MATLAB. In this paper are discussing differences between measurement and simulation results and determine source systematic error. The magnetic field map is measured using the calibration 3D meter and is used for the reconstruction of material properties of the measured sample.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JA - Elektronika a optoelektronika, elektrotechnika
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GA102%2F09%2F0314" target="_blank" >GA102/09/0314: Studium vlastností metamateriálů a mikrovlnných struktur s využitím šumové spektroskopie a magnetické rezonance</a><br>
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2012
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Proceedings of PIERS 2012 in Kuala Lumpur
ISBN
978-1-934142-20-2
ISSN
—
e-ISSN
—
Počet stran výsledku
5
Strana od-do
385-389
Název nakladatele
The Electromagnetic Academy
Místo vydání
Cambridge
Místo konání akce
Kuala Lumpur
Datum konání akce
27. 3. 2012
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—