The article reports on certain artifacts that emerge during the in vitro diffusion-weighted imaging of physical samples. In this context, the authors analyze the influence of magnetic field inhomogeneity, temperature, or eddy currents and consider procedures to mitigation of the artifacts. A technique reducing the examined spurious effects was designed, experimentally verified, and denominated as the three measurements method. The technique proved to be useful mainly for the evaluation of a DWI image measured with a diffusion gradient in the z axis, where the relative measurement error decreased to 3.38% (during measurement using two images, the relative error was greater than 19%). For small errors within the measurement of diffusion constants of a deionised water sample (<5%) it is necessary to select a b-factor value larger than 200x10^6 s.m-2. Temperature stabilisation with an accuracy better than 0.1 degree of celsius during the entire measuring process is a necessary prerequisite for the measurement of biological or material samples with a relative accuracy lower than 1%.
Název v anglickém jazyce
Diffusion MRI: magnetic field inhomogeneities mitigation
Popis výsledku anglicky
The article reports on certain artifacts that emerge during the in vitro diffusion-weighted imaging of physical samples. In this context, the authors analyze the influence of magnetic field inhomogeneity, temperature, or eddy currents and consider procedures to mitigation of the artifacts. A technique reducing the examined spurious effects was designed, experimentally verified, and denominated as the three measurements method. The technique proved to be useful mainly for the evaluation of a DWI image measured with a diffusion gradient in the z axis, where the relative measurement error decreased to 3.38% (during measurement using two images, the relative error was greater than 19%). For small errors within the measurement of diffusion constants of a deionised water sample (<5%) it is necessary to select a b-factor value larger than 200x10^6 s.m-2. Temperature stabilisation with an accuracy better than 0.1 degree of celsius during the entire measuring process is a necessary prerequisite for the measurement of biological or material samples with a relative accuracy lower than 1%.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
JA - Elektronika a optoelektronika, elektrotechnika
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2012
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Measurement Science Review
ISSN
1335-8871
e-ISSN
—
Svazek periodika
12
Číslo periodika v rámci svazku
5
Stát vydavatele periodika
SK - Slovenská republika
Počet stran výsledku
8
Strana od-do
205-212
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—