Diffusion MRI: Mitigation of Magnetic Field Inhomogeneities
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68081731%3A_____%2F12%3A00385190" target="_blank" >RIV/68081731:_____/12:00385190 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.2478/v10048-012-0031-8" target="_blank" >http://dx.doi.org/10.2478/v10048-012-0031-8</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.2478/v10048-012-0031-8" target="_blank" >10.2478/v10048-012-0031-8</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Diffusion MRI: Mitigation of Magnetic Field Inhomogeneities
Popis výsledku v původním jazyce
The article reports on certain artifacts that emerge during the in vitro diffusion-weighted imaging of physical samples. In this context, the authors analyze the influence of magnetic field inhomogeneity, temperature, or eddy currents and consider artifact mitigation procedures. A technique reducing the examined spurious effects was designed, experimentally verified, and denominated as the three measurement method. The technique proved to be useful mainly for the evaluation of a DWI image measured witha diffusion gradient in the z axis, where the relative measurement error decreased to 3.38 % (during measurement using two images, the relative error was greater than 19 %). For small errors within the measurement of diffusion constants of a deionized water sample (< 5 %) it was necessary to select a b-factor value larger than 200106 s.m-2. Temperature stabilization with accuracy better than 0.1 °C during the entire measuring process is a necessary prerequisite for the measurement of bio
Název v anglickém jazyce
Diffusion MRI: Mitigation of Magnetic Field Inhomogeneities
Popis výsledku anglicky
The article reports on certain artifacts that emerge during the in vitro diffusion-weighted imaging of physical samples. In this context, the authors analyze the influence of magnetic field inhomogeneity, temperature, or eddy currents and consider artifact mitigation procedures. A technique reducing the examined spurious effects was designed, experimentally verified, and denominated as the three measurement method. The technique proved to be useful mainly for the evaluation of a DWI image measured witha diffusion gradient in the z axis, where the relative measurement error decreased to 3.38 % (during measurement using two images, the relative error was greater than 19 %). For small errors within the measurement of diffusion constants of a deionized water sample (< 5 %) it was necessary to select a b-factor value larger than 200106 s.m-2. Temperature stabilization with accuracy better than 0.1 °C during the entire measuring process is a necessary prerequisite for the measurement of bio
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
JA - Elektronika a optoelektronika, elektrotechnika
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2012
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Measurement Science Review
ISSN
1335-8871
e-ISSN
—
Svazek periodika
12
Číslo periodika v rámci svazku
5
Stát vydavatele periodika
SK - Slovenská republika
Počet stran výsledku
8
Strana od-do
205-212
Kód UT WoS článku
000314457800007
EID výsledku v databázi Scopus
—