Uncertainty determination in measurements using a Gerdien Tube
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26220%2F14%3APU112338" target="_blank" >RIV/00216305:26220/14:PU112338 - isvavai.cz</a>
Výsledek na webu
<a href="http://piers.org/piersproceedings/download.php?file=cGllcnMyMDE0R3Vhbmd6aG91fDNQMF8xODkwLnBkZnwxNDAzMTkxMTE4MDI=" target="_blank" >http://piers.org/piersproceedings/download.php?file=cGllcnMyMDE0R3Vhbmd6aG91fDNQMF8xODkwLnBkZnwxNDAzMTkxMTE4MDI=</a>
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Uncertainty determination in measurements using a Gerdien Tube
Popis výsledku v původním jazyce
Light negative air ions demonstrably exhibit positive influence on the human organism. The concentration of these ions can be effectively used as the integral criterion for the evaluation of air purity, and there also exists a multitude of other applications. The presented aspiration method enables us to establish the concentration of air ions in the atmosphere. In a Gerdien tube fitted with a one-piece inner electrode, measurement of the saturation characteristics must be performed to facilitate the determination of the air ion mobility spectrum. This method is very sensitivity to noise and leakage current. In this paper is determination uncertainty of type B in Gerdien tube and electrometric amplifier. This uncertainty dependent to input bias current and input noise current used operational amplifier. The next significant source uncertainty type of B is temperature and time unstability quality Teflon insulator. The uncertainty of type A was tested in reference measurement in shielding camber. The result this paper is determine main source of systematic error and estimation uncertainty of type C for our Gerdien Tube.
Název v anglickém jazyce
Uncertainty determination in measurements using a Gerdien Tube
Popis výsledku anglicky
Light negative air ions demonstrably exhibit positive influence on the human organism. The concentration of these ions can be effectively used as the integral criterion for the evaluation of air purity, and there also exists a multitude of other applications. The presented aspiration method enables us to establish the concentration of air ions in the atmosphere. In a Gerdien tube fitted with a one-piece inner electrode, measurement of the saturation characteristics must be performed to facilitate the determination of the air ion mobility spectrum. This method is very sensitivity to noise and leakage current. In this paper is determination uncertainty of type B in Gerdien tube and electrometric amplifier. This uncertainty dependent to input bias current and input noise current used operational amplifier. The next significant source uncertainty type of B is temperature and time unstability quality Teflon insulator. The uncertainty of type A was tested in reference measurement in shielding camber. The result this paper is determine main source of systematic error and estimation uncertainty of type C for our Gerdien Tube.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
—
OECD FORD obor
20201 - Electrical and electronic engineering
Návaznosti výsledku
Projekt
—
Návaznosti
O - Projekt operacniho programu
Ostatní
Rok uplatnění
2014
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
PIERS 2014 Guangzhou Proceedings
ISBN
978-1-934142-28-8
ISSN
1559-9450
e-ISSN
—
Počet stran výsledku
4
Strana od-do
1902-1906
Název nakladatele
The Electromagnetics Academy
Místo vydání
777 Concord Avenue, Suite 207 Cambridge, MA 0213
Místo konání akce
Guangzhou
Datum konání akce
25. 8. 2014
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
000393225900422