EXPERIMENTAL USE OF NOVEL PULSE GENERATOR FOR IRREVERSIBLE ELECTROPORATION
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26220%2F17%3APU123515" target="_blank" >RIV/00216305:26220/17:PU123515 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
EXPERIMENTAL USE OF NOVEL PULSE GENERATOR FOR IRREVERSIBLE ELECTROPORATION
Popis výsledku v původním jazyce
Irreversible electroporation (IRE) is a relatively new technique for minimal invasive nonthermal ablation treatment of solid tumors. This article describes basics of IRE and available devices for this technique. The parameters of commercial NanoKnife IRE generator are verified and the unique experimental IRE generator developed at the Brno University of Technology is presented in this paper. Further, there is a comparison of thermal effects on a target tissue of commonly used radiofrequency ablation (RFA) and the novel IRE generator using an infrared camera. All experiments were performed ex vivo on a porcine liver. The difference in maximum temperature between the output powers used was 17.8 °C with RFA and 3 °C with IRE. This is a significant difference and shows the IRE’s capability of liver tissue ablation without thermal damage. The new IRE generator is also capable of replacing NanoKnife because of its features and versatility.
Název v anglickém jazyce
EXPERIMENTAL USE OF NOVEL PULSE GENERATOR FOR IRREVERSIBLE ELECTROPORATION
Popis výsledku anglicky
Irreversible electroporation (IRE) is a relatively new technique for minimal invasive nonthermal ablation treatment of solid tumors. This article describes basics of IRE and available devices for this technique. The parameters of commercial NanoKnife IRE generator are verified and the unique experimental IRE generator developed at the Brno University of Technology is presented in this paper. Further, there is a comparison of thermal effects on a target tissue of commonly used radiofrequency ablation (RFA) and the novel IRE generator using an infrared camera. All experiments were performed ex vivo on a porcine liver. The difference in maximum temperature between the output powers used was 17.8 °C with RFA and 3 °C with IRE. This is a significant difference and shows the IRE’s capability of liver tissue ablation without thermal damage. The new IRE generator is also capable of replacing NanoKnife because of its features and versatility.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
—
OECD FORD obor
20201 - Electrical and electronic engineering
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2017
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Proceedings of the 23nd Conference STUDENT EEICT 2017
ISBN
978-80-214-5496-5
ISSN
—
e-ISSN
—
Počet stran výsledku
5
Strana od-do
568-572
Název nakladatele
Vysoké učení technické v Brně, Fakulta elektrotechniky a komunikačních
Místo vydání
první
Místo konání akce
Brno
Datum konání akce
27. 4. 2017
Typ akce podle státní příslušnosti
CST - Celostátní akce
Kód UT WoS článku
—